Freeze-dryer Epsilon 1-4 LSCplus
Freeze-dryer Epsilon 2-4 LSCplus
5 Set-up and connection
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
49 / 152
Translation of the original operating manual
Pos : 150 /200 Chris t/361 GT-BA Labor-Pil ot ( PROJ EKTE)/ Epsilon 4er/ Epsilon 1-4_2- 4 LSC plus Standar d/ 050 Aufst ellung und Ansc hl uss/ 050- 0020- 0050-0020 Vak uumpumpe E1-4_2- 4 (Troc kenläuf er) @ 85\m od_1632728134833_68.docx @ 898161 @ 4 @ 1
5. 4. 6
Vak uumpumpe
→
5.2.3.2
Dry-running vacuum pump with silencer
WARNING
Danger of poisoning/infection due to exhaust air
The exhaust air may contain harmful substances originating from the
product.
Contact with particles in the exhaust air (especially by inhalation) may
result in severe damage to health.
• Use a silencer with integrated filter elements or – depending on the
product
– guide the exhaust gases from the outlet of the vacuum pump
to a suitable treatment system in order to prevent the discharge of
harmful substances into the surrounding atmosphere! Compliance with
the national rules and regulations for the protection of the environment
must be ensured!
• When working on the vacuum system (and especially on the vacuum
pump), it may be necessary to wear suitable protective clothing,
gloves and a breathing mask depending on the product!
NOTE
Refer to the separate instruction manual of the vacuum pump and
silencer!
The vacuum pump must be connected to the vacuum connection of the unit
and to the electrical socket at the back of the unit, which is marked
accordingly (see chapter 2.1.1 - "Functional and operating elements").
NOTE
The vacuum pump is supplied with power by the unit, but the maximum
current for the vacuum pump is limited. It is absolutely essential to refer to
the label of the electrical outlet for the vacuum pump (see the following
picture)!
If the current requirement of the vacuum pump is higher than the value
that is stated on the label, the pump must be supplied separately via an
on-site power socket.
1
Label indicating the
maximum current
Fig. 13: Indication of the maximum current for the vacuum pump (example)