Appendix D - Essential performance
The performance characteristics essential to the correct operation of the
Primo Combination 860 are:
Power input
100-240V ac
Treatment time
0 – 30 minutes
Ultrasound
Frequency
1.1 MHz ±5% and 3.4 MHz ±5%
Maximum intensity
1.5 W/cm
2
(±20%) in CW
3.0 W/cm
2
(±20%) pulsed modes
Maximum output power
6 W (±20%) average
Output modes
CW and pulsed 1:1, 1:2, 1:4 and 1:9
Pulse duration
2 ms (±10%)
Large transducer
ERA 4cm
2
(±20%) (IEC61689), collimated
BNR<5
Small transducer
ERA 0.6cm
2
(±20%) (IEC61689),
divergent at 1MHz
ERA 0.4cm
2
(±20%) (IEC61689),
collimated at 3MHz
BNR<5
Interferential 4-pole
Carrier frequency
2 kHz, 4 kHz or 8 kHz ) (± 5%)
AMF
0 – 250 Hz in 1 Hz increments ) (±2%)
Swing pattern
1|1, 6|6 or 6^6
Output type
CC
0-100mA peak (±10%) into 500
Ω
CV
0-70V peak (±10%) into 2 k
Ω
Interferential 2-pole
Carrier frequency
2 kHz, 4 kHz or 8 kHz (± 5%)
AMF
0 – 250 Hz in 1 Hz increments (±2%)
Swing pattern
1|1, 6|6 or 6^6
Output type
CC
0-100mA peak (±10%) into 500
Ω
CV
0-70V peak (±10%) into 2 k
Ω
Russian Stimulation
Carrier frequency
2.5 kHz (± 5%)
Modulation frequency
1 – 100 Hz (± 5%)
Surges
1:1 to 1:5
Output type
CC
0-100mA peak (±10%) into 500
Ω
CV
0-70V peak (±10%) into 2 k
Ω