Combination 860
23
If the transducer face is lifted from the treatment site or if for any reason
there is insufficient contact between the transducer and the treatment site
for more than two seconds, the power applied to the transducer will also be
reduced to a low level. The treatment light on the transducer will turn off,
the treatment time will cease to count down and the status bar will display
CONTACT, indicating that the required output cannot be delivered. An
audible alarm will sound if this option has been selected in the Setup menu.
When good contact is restored, the treatment indicator on the transducer
will light, the status bar will display TREATMENT and the timer will continue
to count down.
If the output intensity is returned to zero using the rotary control, before the
treatment time has elapsed, the display will show the treatment time
remaining. When the intensity is increased again the treatment will
continue.
When the treatment time reaches 00:00, treatment is terminated. The
intensity and power displays will go to zero, ultrasonic power from the
transducer will be turned off, the treatment indicator will turn off and the unit
will give a two second beep. Remove the transducer from the treatment
site, wipe off any coupling medium and return the transducer to its cradle
on the front of the unit.
Remove the remaining coupling medium from the treatment site.
The transducers are also suitable for treatment using a water bath. This is
especially useful when treating areas which are not uniform such as feet or
hands. When using a water bath it is advisable to use degassed water
(water that has been boiled to remove any air and then allowed to cool).
After the part of the body has been immersed in the water, remove any air
bubbles that may have accumulated on the skin. Set up the treatment
parameters and then immerse the transducer in the water before turning
the output on. Hold the transducer with its face approximately 1 cm away
from the treatment site and using the rotary control set the required
intensity remembering to keep the transducer moving in small circular
paths to prevent standing waves. At the end of the treatment the intensity
and power displays will read zero, and the ultrasound power will turn off.
Remove the transducer from the water and dry both it and the area treated.