Combination 860
64
Diadynamic
In diadynamic mode the unit produces six different waveforms.
The maximum peak output current is limited to 70 mA.
DF - diaphasé fixe
The DF waveform is a continuous full wave rectified 50 Hz sinewave.
rms current = peak current x 0.707
The maximum rms current is 50 mA.
MF - monophasé fixe
The MF waveform is a continuous half wave rectified 50 Hz sinewave.
rms current = peak current x 0.5
The maximum rms current is 37.5 mA.
CP - modulé en courtes périodes
The CP waveform is a combination of the MF and DF waveforms. The unit
provides 1 s of MF (half wave signal) followed by 1 s of DF (full wave
signal), the sequence being repeated continuously.
CPiso - modulé en courtes périodes isodynamique
This is the same as the CP waveform except that the amplitude of the MF
signal is 12.5% less than the amplitude of the DF signal.