Electrode error detection
When a constant current output is chosen and the unit is operating in 4 or 2
pole interferential, russian, diadynamic, sinusoidal, galvanic or träbert
modes, the electrode impedance is monitored to ensure that adequate
electrode contact is maintained.
If the unit detects an electrode impedance too high to safely deliver the
required current, then the output of the Combination 860 is terminated, an
error message is displayed on the bottom line of the LCD and an
intermittent alarm is sounded.
To cancel the alarm and clear the error message, press any button or turn
the rotary control anticlockwise. The remaining treatment time is
maintained. Check the electrodes and leads before continuing treatment. If
rubber pad electrodes and sponge covers are being used check that they
are held securely with even pressure by the elasticated bandages and that
the sponges have not dried out.