The following describes the set-up pages for each stimulation type when
accessed by pressing the relevant button in the stimulation screen:-
4-pole interferential set-up
Treatment time:
Is selected either by touching the digits of the time display
or by pressing the clock symbol and entering the desired treatment time.
The time can be set in 30s intervals.
Carrier:
Pressing this button selects 4, 8 or 2 kHz carrier frequency.
AMF:
Pressing this button opens the AMF window.
The amplitude modulation frequency (AMF), or beat frequency, is set as a
Base and Peak beat frequency. The beat frequency sweeps between the
base and the peak frequency at a rate determined by the setting of the
Pattern button. If the base and peak frequencies are set to the same value
then a constant beat frequency is produced. The Base and Peak
frequencies may be set in 1 Hz increments from 0 to 250 Hz. Press OK
when the desired frequencies have been chosen to return to the 4-pole
Interferential set-up screen.