Galvanic
Galvanic mode produces a direct current from 0 to 70 mA in either a
positive, negative or auto-reverse (half way through the treatment time)
electrical polarity. Two channels of output are available with independent
intensity control.
Interrupted galvanic
Interrupted galvanic mode produces three standard pulse shapes.
Rectangular pulses are available from 10 µs to 1 s pulse width and
triangular and trapezoidal pulses from 1 ms to 1s. The pulse repetition rate
is from 2 to 30 pulses per minute. Two channels of output are available with
independent intensity control, and their polarity may be selected as Pos (+),
Neg (-), or autorev (half way through the treatment time).
Träbert
This mode produces a continuous train of 2ms pulses with a 5 ms interval
between each pulse. The pulse repetition rate is therefore approximately
143 Hz. Two channels of output are available with independent intensity
control. Pos (+), Neg (-), or autorev polarities are selectable.