INST.No.INE-454-P2 Software
Ver.1.2
- 4 -
1) To program the reference temperature value, press
SEL
key in the measurement mode screen until “
Auto
”
appears in the sub display
as shown in the right figure.
2) By pressing
key, the most significant digit of the main
display will blink.
3) Use
or
key to program the desired value.
4) To shift to the next lower digit, press
key.
The next lower digit will blink.
5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
6) To activate this programming, press
ENT
key.
1) To program the emissivity value, press
SEL
key in the
measurement mode screen until “
ε
” appears in the sub
marker
as shown in the right figure.
By pressing
SEL
key Indicate the 1 wavelength selection
display under the main marker.
2) By pressing
key, the most significant digit of the sub
display will blink.
3) Use
or
key to program the desired value.
4) To shift to the next lower digit, press
key. The next
lower digit will blink.
5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
6) To activate this programming, press
ENT
key.
3.1.2 Emissivity (ratio) programming 2
: Program the emissivity (ratio) of 1.35/0.9
µ
m in
“2col” and 0.9
µ
m in “1col” in case of automatic
emissivity calculation.
In case of automatic emissivity calculation, program the emissivity ratio at high temperature side range.
Even if it sets up in case of 2 color measurement and single color measurement it is not related to
measurement.
3.2 Automatic emissivity calculation
The emissivity is automatically calculated when you coincident the measured value to the reference
temperature value entered by function keys or by an analog input (option).
3.2.1 Automatic emissivity calculation(Internal ):Input of reference temperature
In the measurement mode, the emissivity is automatically calculated by entering the reference
temperature value measured by a thermocouple or other sensor in advance.
This function is effective when “
In
” (internal) or “
Ext
” (external) is selected in
[4.7 Automatic emissivity calculation type selection.].
R e m a r k s
•
The programmable range is
0
to
6280ºC
.
•
The default is the minimum temperature value of the measuring range.
R e m a r k s
3. Operating mode
Tb CONT MEM PEAK
AL
AH
C
F
ε
Sub display
Sub marker
Main display
1 indicated
The programmable range is
1.999
to
0.050
.
The default value is
1.000.
R e m a r k s
AL
AH
C
F
ε
Tb CONT MEM PEAK
Sub display
Main display