INST.No.INE-454-P2 Software
Ver.1.2
-24-
4.11 Analog input programming
(
Model IR-CAQ
5
only
)
4.11.1 Selection of analog input items
This mode is for the selection of the analog input (analog input disable, for remote emissivity
programming or for automatic emissivity calculation).
1) To select the analog input item, press
SEL
key in the
engineering mode screen until “
Ain
” appears in the sub
display
as shown in the right figure.
2) By pressing key, “
no
”, “
ErMt
” or “
EAut
” will blink in
the main display.
3) Use or key to select the desired type.
4) To activate this selection, press
ENT
key.
[Analog input items]
no
: Analog input disable
ErMt
: For remote emissivity
programming
EAut
: For automatic emissivity
calculation
Tb CONT MEM PEAK
AL
AH
C
F
ε
[Analog input disable]
Main display
Sub display
Tb CONT MEM PEAK
AL
AH
C
F
ε
[ Remote emissivity
programming]
Main display
Sub display
Tb CONT MEM PEAK
AL
AH
C
F
ε
[ Automatic emissivity
calculation]
Main display
Sub display
•
The default is
no
(analog input disable).
R e m a r k s
4. Engineering mode