16-3
4. Programming flow chart
<Example> Adjustment of channel 1 (printing range -50 to + 150)
Operation screen
Apply an input equivalent to the lower-limit value by
a standard tool.
Apply an input equivalent to the higher-limit value
by a standard tool.
Stored
[Engineering mode]
[Selecting adjustment]
[Selecting measured
value]
[Programming channel]
Cursor
Cursor
Adjustment
Measured value
Cursor
Channel
Cursor
[To zero adjustment
display]
(Zero) Lower-limit value
[Zero adjustment]
[To span adjustment
display]
(Span) Higher-limit value
[Span adjustment]
[To programming
channel display]
Channel
Cursor
( + ) Longer
than 3sec.
Longer
than
2sec.
+
Remarks 1
Remarks 2
Remarks 3
Reference
(Calculation of zero
correction data)
(Calculation of span
correction data)
Calculated correction data are canceled
by returning to the operation display at
the procedure before [storing].
For returning to operation
screen
Remarks 1
When a skipped channel is
programmed, it does not accept the
[ENTRY] key.
Skipped channel does not
accept any [ENTRY] key
Remarks 2
After reconnecting to input terminals
to be adjusted, repeat the same
procedure line from the programming
channel display (indicated by a dotted
line).
Other channels adjustment
Remarks 3
By pressing [ENTRY] key after clearing
(by pressing + keys) in zero
adjustment or span adjustment display,
the correction data of the channel
becomes the initial value.
Initialization of correction data
Remarks 3
Store the calculated correction data into
memory.
Storing
Reference
Содержание AL3000 SERIES
Страница 1: ...INST No INE 270J AL3000 SERIES MULTI POINT TYPE HYBRID RECORDER ...
Страница 2: ......
Страница 25: ...MEMO ...
Страница 114: ...MEMO ...
Страница 115: ......
Страница 116: ...32 8 KUMANO CHO ITABASHI KU TOKYO 173 8632 Telephone 81 3 3956 2171 Facsimile 81 3 3956 0915 Printed in Japan ...