7-2
7.2. Default Parameters/Functions
and Programming Ranges
1.Parameters to be programmed before operation
Parameters
Programming
items
Defaults Programming
ranges
Entry
(Note)
Range No.
07 (
±
5V)
01 to 10 (Voltage), 21 to 56 (Thermocouple),
70 to 80 (Resistance thermometer)
RJ enabled/
disabled
0 (Disabled)
0 (Disabled),
1 (Enabled, Thermocouple range only)
Range/
printing
range
Printing range 0.00 to 5.00
Max. 10 digits (excluding decimal point)
+
°C /
°
F
operation
selection
- °C
operation
°C, °F operation selection
(thermocouple and resistance thermometer
ranges only)
+
Chart speed
-
20(mm/h)
0001 to 1500(mm/h), 1mm step
Time
-
Japanese time
1998. 01. 01. 00:00 to 2097. 12. 31. 23:59
2. Other programmable parameters and functions
Parameters
Programming
items
Defaults
Programming ranges
Entry (Note)
Scale
-
0.00 to 5.00
Printing range reflects on the scale.
Accordingly, no programming is necessary
for thermocouple and resistance
thermometer ranges, except for a change of
decimal point.
Max. 11 digits (excluding decimal point)
Lower-limit value*
Higher-limit value*
*Within 5 digits
+
Skip
(deletion and
recovery of
channel)
-
Not
programmed
In the programming of [Range/printing
range]
Channel to be deleted
→
Clear the range/printing range.
Channel to be recovered
→
Program the range/printing range.
+
Subtract
printing
(For printing of
difference)
-
Not
programmed
In the programming of [Range/printing
range]
Subtract printing channel between channels
= Reference channel - subtraction channel
Subtract printing channel with reference
value = Reference channel - reference value*
*Put a decimal point to one place of
decimals. (within 5 digits excluding decimal
point)
+
Channel
1 to n, n: Number of channels = 6
Level
Not
programmed
1 to 4
Absolute
value
H (Higher-limit),
L (Lower-limit)
Change-of-
ratio
U (Increase-limit),
d (Decrease-limit)
Alarm
types
H
Defferential
b (Differential higher-limit),
S (Differential lower-limit)
Alarms
1.Absolute value
alarm
2.Rate-of-
change alarm
3.Differential
alarm
Output
number
0
[0] fixed…Program at optional alarm output
only
+