background image

Page 16 

5.7 ‘Advanced’ Page 

The advanced page contains functions that will change the how the microDAQ acts  and how it 
applies various calculations to its data, this page should only be used by users who fully 
understand what they are changing. 
 

5.7.1  Filtering 

The microDAQ has settings to allow the user flexible control over the data throughput of the device.  
Averaging  options  allow  the  preference  of  noise  reduction  over  time  domain  response,  the 
frequency of calibration temperature compensation renewal may be chosen as may be the size of 
the  average  (and  hence  time  taken)  for  the  rezero  routine.  Table  5.2  details  the  function  of  the 
signal parameter option controls. 
 

 

Figure 5.9, Filtering group 

 

 

Control 

Function 

'Pressure Input Impulse Filter' 

Check box to apply impulse filter to pre calibration data - will remove 
single impulse noise events in the pressure data. 

'Temperature Input Impulse Filter' 

As above but with temperature data. 

'Pressure input average samples' 

Selects the number of samples for a moving average of pre calibration 
pressure data. 

'Temperature input average samples' 

As above but with temperature data. 

'Temperature Compensation' 

Selects the temperature compensation scheme for the calibration.  
'Continuous' repeatedly rebuilds the calibration data on a channel by 
channel basis without interrupting the flow of data.  'With zero only' 
rebuilds the table when a user rezero is issued, after the rezero has 
been actioned. 

‘Apply’ 

Applies the settings to the local settings memory 

Table 5.8, Filtering group settings 

 

5.7.2  Advanced communication settings 

The microDAQ has several more advanced communication options available, but these features 
are unlikely to be used by an everyday user and as such should generally be left as default 
settings. 

 

Figure 5.10, Advanced communication settings 

Содержание microDAQ-Mk2

Страница 1: ...Road North Walsham Norfolk NR28 0AJ ENGLAND Tel 01692 500555 Fax 01692 500088 microDAQ Mk2 Pressure Scanner Acquisition System INSTALLATION AND OPERATING MANUAL e mail info chell co uk Visit the Chell...

Страница 2: ...ied in this manual may impair the user s protection Chell Document No 900193 Issue 1 3 ECO 2230 Date 9th May 2016 Chell s policy of continuously updating and improving products means that this manual...

Страница 3: ...erver 8 5 1 Introduction 8 5 2 Common Controls Sidebar 9 5 3 The Setup Page 9 5 3 1 Introduction 9 5 3 2 Data Streaming 9 5 3 3 TCP Parameters 10 5 3 4 UDP Parameters 10 5 3 5 CAN Parameters 11 5 4 Li...

Страница 4: ...date both conventional HD series scanners and Digital Thermal Compensation DTC scanners with 16 32 and 64 channels The microDAQ is supplied with all the software required to configure it and read the...

Страница 5: ...ons USB2 0 micro 57600 baud no parity 8 data bits and 1 stop bit 2 4 Ethernet Specifications TCP IP 10Mb s 100Mb s via Auto Negotiation TCP UDP protocols supported 2 5 Operating conditions Operating t...

Страница 6: ...ge to the unit Always switch the power at the power supply source 3 3 Scanner Connector Internal micro miniature D type female Pin number Signal 1 Address line A0 2 Address line A1 3 Address line A2 4...

Страница 7: ...atus If there isn t a valid connection the process waits until either a network is found or the timeout occurs This timeout is configurable via setup and ranges from 0 to 30 secs per check 2 checks Wh...

Страница 8: ...d set and so on When the hardware trigger is activated the microDAQ will wait for the first pulse The time that this first pulse is generated can be measured by the user and therefore the time of the...

Страница 9: ...s the tools for performing a new calibration namely calculating and downloading new coefficients and the means to managing calibration data Data from the calibration procedure and resulting coefficien...

Страница 10: ...ach category is detailed separately in the following 5 3 2 Data Streaming The Data Streaming section allows the user to change settings that affect all three communication protocols and allows the use...

Страница 11: ...ory Table 5 3 TCP Comms group settings 5 3 4 UDP Parameters The UDP section figure 5 5 holds all the settings specific to UDP In UDP mode each acquisition cycle of x number of channels is packed as a...

Страница 12: ...whether an acknowledgement of these user commands is sent on the next higher message number Data may be transmitted on either multiple messages or alternatively on a single message ID with a selectab...

Страница 13: ...16 bit unsigned calibrated decimal 16 bit unsigned and engineering units or voltage scaled to the known full scales Also for DTC scanners the temperature values as raw unsigned 16 bit and calibrated e...

Страница 14: ...non temperature calibration may be performed at a single temperature Figure 5 9 shows the Calibration tab controls which are further explained in table 5 8 The Live values section is an indicator of...

Страница 15: ...emperature calibration Unchecked only temperature point 1 is available to the user and the system will calculate coefficients for a non temperature calibration Live Values The live values of the curre...

Страница 16: ...rol Function Range Shows the floating point value for the range contained within the scanner x3 derange Shows the floating point value for the sensitivity derange constant contained within the scanner...

Страница 17: ...n data will remove single impulse noise events in the pressure data Temperature Input Impulse Filter As above but with temperature data Pressure input average samples Selects the number of samples for...

Страница 18: ...TSEG1 TSEG2 SJW Register values for the CAN module within microDAQs microcontroller Table 5 9 Advanced communication settings 5 7 3 Miscellaneous The remaining parameters are edited via the Miscellan...

Страница 19: ...AQ s internal calibration data table If I daq detected use raw calibration If checked this replaces the normal polynomial calibration function with a raw linear function whenever an I daq is present i...

Страница 20: ...ear calibration Figure 5 12 Span and zero coefficients 5 9 Timestamp page This page allows the user to edit the timestamp settings of the microDAQ This timestamp will alow the user to get millisecond...

Страница 21: ...the timestamp for every channel It should be noted that the latter 2 options will reduce the maximum transmission speed datastream get PC time This button allows the user to get the timestamp from the...

Страница 22: ...ual basis and Chell Instruments Ltd Provides a fully traceable facility for this purpose 6 3 Adjustment There are no user adjustments in the instrument The user is strictly forbidden from removing the...

Страница 23: ...uter Gnd 7 3 Can Connector Mating connector 9 way dtype male Pin Number Designation 2 CAN_L 5 Gnd 7 CAN_H 4 N C 5 0V 7 4 Trigger Connector Mating connector 50 Ohm BNC Pin Number Designation Center TTL...

Отзывы: