5.4. 'Live Data' Page
Figure 5.8 shows the 'Live Data' page of the webserver, selected for a 64 channel pressure
scanner.
Figure 5.6, Live Data Page
The live data page is a means to demonstrating the correct operation of microDAQ and testing the
unit's calibration. By selecting the correct number of channels for the attached scanner, the
software shows a data label dedicated to each. The type of value shown in the label may be
selected by means of the option buttons to the right of the frame, the user selecting between raw
(decimal 16 bit unsigned), calibrated (decimal 16 bit unsigned) and engineering units or voltage
scaled to the known full scales. Also for DTC scanners, the temperature values as raw unsigned
16 bit and calibrated engineering units may be displayed. Values are updated automatically, once
a second, with the default view being Pressure (Eng). Use the Select button to start showing values
from one of the other selectable options.
The 'Derange' shows whether the derage is active or inactive, this is changed by changing the DTC
Gain value in the advanced page, a DTC Gain of 1 means derange will be inactive and a DTC Gain
of 3 means derange will be active. The derange value will be displayed on if derange is active, if
not then the displayed value will be 0.