background image

Page 14 

 

Control 

Function 

Temperature tickbox and radio 
buttons 

Select at which temperature point the unit is currently been calibrated 

Applied Pressure 

The engineering units pressure setpoints - user entered, or set to standard values (+/- 
FS, +/-0.5FS, 0) by clicking 'Defaults' 

Pressure (16 bit FS) 

The 16 bit unsigned representation of the pressure setpoints, calculated by the 
software 

Read value (16 bit FS) 

The raw value read by microDAQ from the pressure scanner for the pressure setpoint 

Temperature (16 bit FS) 

The temperature value read by microDAQ from the pressure scanner at that pressure 
setpoint 

‘Accept’ button 

Clicking 'Accept' stores the current read value as the calibration point, and moves on 
to the following setpoint 

‘Defaults’ button 

Sets the default calibration points as +/- full scale, +/- 0.5 full scale and zero 

Calculate 

Takes the acquired data and calculates the derived calibration coefficients - will not 
proceed unless there is no more than one zero in the applied pressure text boxes 

Clear 

Clear the current calibration test data to zeroes, also clears the calibration controls 

Temperature Cal. 

When checked enables temperature options  2 and 3 for a 3 temperature calibration.  
Unchecked only temperature point 1 is available to the user and the system will 
calculate coefficients for a non temperature calibration 

Live Values 

The live values of the current channel are displayed here 

Live update on/off 

Selects whether the live values will update at a 1 second interval or not. 

View channel 

Select the channel that supplies the values displayed in the calibration controls 

C0 – C8 

Calibration coefficients read/calculated for the selected channel 

Burn Coeffs 

Tell the microDAQ to store the current coefficients to non-volatile storage (EEPROM) 

Linear Cal Functions 

‘Set Span’ 

Set the span of a linear calibration – confirms the value of pressure that has been set 
up as the span value is being applied to the sensor. 

‘Reset Linear Cal.’ 

Resets the linear calibration to (+ 0) x1 i.e. no offset, unity gain.  Applies ONLY to the 
current linear calibration, ie the DTC calibration if using a DTC scanner. 

Calibration Load/Save 

Browse 

Allows the user to search through the file directory system and select a calibration file  

Save File 

Save the current local calibration coefficients to a file - defaults to the calibration file 
directory 

Read File 

Reads the calibration file selected by the browse function into the microDAQ (note: 
Reading a non calibration file will cause the system to crash) 

 

Table 5.6, Calibration Page controls 

 
 
 

Содержание microDAQ-Mk2

Страница 1: ...Road North Walsham Norfolk NR28 0AJ ENGLAND Tel 01692 500555 Fax 01692 500088 microDAQ Mk2 Pressure Scanner Acquisition System INSTALLATION AND OPERATING MANUAL e mail info chell co uk Visit the Chell...

Страница 2: ...ied in this manual may impair the user s protection Chell Document No 900193 Issue 1 3 ECO 2230 Date 9th May 2016 Chell s policy of continuously updating and improving products means that this manual...

Страница 3: ...erver 8 5 1 Introduction 8 5 2 Common Controls Sidebar 9 5 3 The Setup Page 9 5 3 1 Introduction 9 5 3 2 Data Streaming 9 5 3 3 TCP Parameters 10 5 3 4 UDP Parameters 10 5 3 5 CAN Parameters 11 5 4 Li...

Страница 4: ...date both conventional HD series scanners and Digital Thermal Compensation DTC scanners with 16 32 and 64 channels The microDAQ is supplied with all the software required to configure it and read the...

Страница 5: ...ons USB2 0 micro 57600 baud no parity 8 data bits and 1 stop bit 2 4 Ethernet Specifications TCP IP 10Mb s 100Mb s via Auto Negotiation TCP UDP protocols supported 2 5 Operating conditions Operating t...

Страница 6: ...ge to the unit Always switch the power at the power supply source 3 3 Scanner Connector Internal micro miniature D type female Pin number Signal 1 Address line A0 2 Address line A1 3 Address line A2 4...

Страница 7: ...atus If there isn t a valid connection the process waits until either a network is found or the timeout occurs This timeout is configurable via setup and ranges from 0 to 30 secs per check 2 checks Wh...

Страница 8: ...d set and so on When the hardware trigger is activated the microDAQ will wait for the first pulse The time that this first pulse is generated can be measured by the user and therefore the time of the...

Страница 9: ...s the tools for performing a new calibration namely calculating and downloading new coefficients and the means to managing calibration data Data from the calibration procedure and resulting coefficien...

Страница 10: ...ach category is detailed separately in the following 5 3 2 Data Streaming The Data Streaming section allows the user to change settings that affect all three communication protocols and allows the use...

Страница 11: ...ory Table 5 3 TCP Comms group settings 5 3 4 UDP Parameters The UDP section figure 5 5 holds all the settings specific to UDP In UDP mode each acquisition cycle of x number of channels is packed as a...

Страница 12: ...whether an acknowledgement of these user commands is sent on the next higher message number Data may be transmitted on either multiple messages or alternatively on a single message ID with a selectab...

Страница 13: ...16 bit unsigned calibrated decimal 16 bit unsigned and engineering units or voltage scaled to the known full scales Also for DTC scanners the temperature values as raw unsigned 16 bit and calibrated e...

Страница 14: ...non temperature calibration may be performed at a single temperature Figure 5 9 shows the Calibration tab controls which are further explained in table 5 8 The Live values section is an indicator of...

Страница 15: ...emperature calibration Unchecked only temperature point 1 is available to the user and the system will calculate coefficients for a non temperature calibration Live Values The live values of the curre...

Страница 16: ...rol Function Range Shows the floating point value for the range contained within the scanner x3 derange Shows the floating point value for the sensitivity derange constant contained within the scanner...

Страница 17: ...n data will remove single impulse noise events in the pressure data Temperature Input Impulse Filter As above but with temperature data Pressure input average samples Selects the number of samples for...

Страница 18: ...TSEG1 TSEG2 SJW Register values for the CAN module within microDAQs microcontroller Table 5 9 Advanced communication settings 5 7 3 Miscellaneous The remaining parameters are edited via the Miscellan...

Страница 19: ...AQ s internal calibration data table If I daq detected use raw calibration If checked this replaces the normal polynomial calibration function with a raw linear function whenever an I daq is present i...

Страница 20: ...ear calibration Figure 5 12 Span and zero coefficients 5 9 Timestamp page This page allows the user to edit the timestamp settings of the microDAQ This timestamp will alow the user to get millisecond...

Страница 21: ...the timestamp for every channel It should be noted that the latter 2 options will reduce the maximum transmission speed datastream get PC time This button allows the user to get the timestamp from the...

Страница 22: ...ual basis and Chell Instruments Ltd Provides a fully traceable facility for this purpose 6 3 Adjustment There are no user adjustments in the instrument The user is strictly forbidden from removing the...

Страница 23: ...uter Gnd 7 3 Can Connector Mating connector 9 way dtype male Pin Number Designation 2 CAN_L 5 Gnd 7 CAN_H 4 N C 5 0V 7 4 Trigger Connector Mating connector 50 Ohm BNC Pin Number Designation Center TTL...

Отзывы: