![Chauvin Arnoux Green TEST-FTV 100 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/chauvin-arnoux/green-test-ftv-100/green-test-ftv-100_user-manual_2594607009.webp)
GREEN TEST-FTV 100
English
Page 9 sur 47
5. PREPARING THE SCREEN
The GREEN TEST FTV 100 has a 5.7" LCD screen that displays all measurements and the results of the
efficiency calculations in real time.
SYSTEM
1
POWER
[
kW
]
:
]
:
7.50
MIN RAD
[
W/m2
]
:
600.00
PANEL EFFICIENCY
--.-%
INVERTER EFFICIENCY
--.-%
ENVIRONMENT MEASUREMENT
Radiation W/m
2
: 0.00
environm temp. °C: ---.--
panel temp. °C: ---.--
temp. coeff.: -.--
THEORATICAL
AVAIL POWER
[
kW
]
:
0.00
DC MEASUREMENT
ERROR
V1: -
∗∗∗∗
-
I1: - 37.9A
V2: -
∗∗∗∗
-
I2: - 38.1A
V3: -
∗∗∗∗
-
I3: - 38.9A
DC POWER
[
kW
]
:
0.00
AC MEASURE
ERROR
V1: -
∗∗∗∗
-
I1: 0.00A
V2: -
∗∗∗∗
-
I2: 0.00A
V3: -
∗∗∗∗
-
I3: 0.00A
AC POWER
[
kW
]
:
0.00
FUNCTION MENU
INSTALLATION
MEASURE
ADJUSTMENT
CONFIGURATION
REMOTE MEASURE
OFF
WARNING
1.
IMP: indicates the number of the photovoltaic installation selected
2.
POWER [kW]: indicates the nominal power of the photovoltaic installation (value entered by the
user).
3.
MIN RAD [W/m²]: indicates the minimum insolation required for the performance calculations
(see ADJUSTMENTS MENU). The default value is 600 W/m².
4.
ENVIRONMENT MEASURES:
Radiation W/m
²
:
indicates the insolation measured by pyranometer (located on the
panels and having the same inclination) (see also the directions
for use of the pyranometer).
temp. environm °C:
indicates the value of environme nt temperature measured by a
Pt100 probe.
temp. panels °C:
indicates the temperature of the solar panels measured by a
Pt100 probe.
temp. coeff.:
indicates the panel performance correction parameter.
This value changes when the temperature is above the data
entered (NOCT).
It is a constant 0.85 when the temperature is between 0°C and
the NOCT (normal operating cell temperature).
1
2
3
4
5
6
7
8
9
11
12
14
13
15
Содержание Green TEST-FTV 100
Страница 1: ... Testing instrument for photovoltaic installations ENGLISH User s manual ...
Страница 14: ...GREEN TEST FTV 100 English Page 14 sur 47 ...
Страница 15: ...GREEN TEST FTV 100 English Page 15 sur 47 ...
Страница 35: ...GREEN TEST FTV 100 English Page 35 sur 47 Example of a Report of on site measurement on the tests performed ...