With this function it is not only possible to test classic
diodes but also LED’s and any other semiconductor where
the junction corresponds to a direct voltage less than 3 V.
■
Protection : same as for Resistance
6. TEMPERATURE
In order to avoid electrocution, never put the extremity of
the K thermocouple in contact with a conductor whose
voltage is greater than 24 V RMS or 60 V ... in relation to
earth.
■
Select the TEMP function.
■
Connect the adapter to the COM and TEMP (K) terminals,
the “+” of the adapter must be connected to the TEMP
terminal (see fig. below).
■
Connect the K thermocouple to the adapter.
■
Automatic range selection: read off the value
■
Select the measurement unit, °C or °F
■
If necessary, store the value in memory by pressing the
HOLD button
If necessary, press the following buttons :
■
MIN MAX to read off the min. or max. values
■
to light up the display
°C
-40°C...0°C
0°C...400°C
400°C...1000°C
Resolution
1°C
Accuracy
±5% R + 2°C ±1% R + 2°C ±3% R + 2°C
Protection
HRC fuse 0.1 A / 660 V
°F
-40°F...+ 150°F 150°F...750°F 750°F...2000°F
Resolution
1°F
Accuracy
±5% R + 4°F ±1% R + 4°F ±3% R + 4°F
Protection
HRC fuse 0.1 A / 660 V
NB :
If there is no K thermocouple at the input, "OL" is
displayed.
COM
TEMP (K)
-
+
+
K
Multimeter
terminals
Adaptor
K thermocouple
Connector