10
To exit from the MAX AVG MIN mode, long-press the
MAX AVG MIN
key.
3.2.3. MAP FUNCTION
The MAP function is used to map the illumination on a surface or in a room.
A long press on the
MAP
key activates the function. The counter of the number of points recorded reads zero.
Plot the map of the room and situate the measurement points.
Here’s an example of 2 rooms.
The represent the luminaires.
The represent the measurement points.
Refer to §4.4 of standard NF EN 12464-1 to establish a map in conformity with good practice.
Place the sensor on the first measurement point and press the
MEM
key to record the value in memory. The counter is incremented.
lx
P
MAP
MAP
lx
P
MIN/AVG/MAX
MAP
MAP
MEM
REC
Proceed in the same way for all the other points of the map.
When all of the points have been entered, you can look up the maximum, the mean, and the minimum of the values entered. To
do this, press the
MAX AVG MIN key 3 times.
To exit from the MAP mode, long-press the
MAP
key.
MAX
AVG
MIN
MAX
AVG
MIN
MAX
AVG
MIN
MAX
AVG
MIN
MAX
AVG
MIN
MAX
lx
Fluo.
Fluo.
lx
lx
Fluo.
AVG
MIN/AVG/MAX
TIME
MIN/AVG/MAX
TIME
MIN/AVG/MAX
TIME
lx
Fluo.
MIN
MIN/AVG/MAX
TIME
P
P
P
P