11
For each map measurement, a file is created with all of the measurement points. This file can then be recovered for analysis.
3.3. RECORDING THE MEASUREMENTS
A short press on the
MEM
key records the measurement and the date.
A long press on the
REC
key starts or stops a recording session.
To see the records, you must use a PC and install the Data Logger Transfer software (see §4).
3.4. ERRORS
The instrument detects errors and displays them in the form Er.XX. The main errors are the following:
Er.01: Hardware malfunction detected. The instrument must be sent in for repair.
Er.02: Error in internal memory. Format it using Windows.
Er.03: Hardware malfunction detected. The instrument must be sent in for repair.
Er.10: The instrument has not been adjusted or is not correctly adjusted. The instrument must be sent back to customer service.
Er.11: The update of the internal software is not compatible with the instrument (the software is that of another instrument of the
line). Install the correct internal software in your instrument.
Er.12: The update of the internal software is not compatible with the electronic boards in the instrument. Reload the previous
internal software into your instrument.
Er.13: Recording scheduling error. Check that the instrument’s time and the time of the Data Logger Transfer software are the same.