Intelect® Advanced Therapy System
104
INSTALLATION/REMOVAL
Remove Screws Securing Module
With a #1 Phillips Screwdriver, remove the
four 4 mm x 20 mm screws securing the
Module to the Therapy System.
REMOVAL CHANNEL 3/4 ELECTROTHERAPY MODULE (continued)
Disconnect Ribbon Cable at Module
Separate the Module from the Therapy
System and disconnect the Ribbon Cable
from the Module Connector Pins.
Do not disconnect Ribbon Cable from
the Therapy System.
DISCONNECT
RIBBON FROM
MODULE ONLY
Store and Secure Ribbon Cable
Roll the Ribbon Cable up and store in
the cavity of the Therapy System. Secure
Ribbon Cable with a nonpermanent
adhesive tape.