Intelect® Advanced Therapy System
SPECIFICATIONS
19
WAVEFORM SPECIFICATIONS (continued)
VMS is a symmetrical biphasic waveform with a 100 µsec interphase
interval. Because the pulse is relatively short, the waveform has a low
skin load, making it suitable for applications requiring high intensities,
such as in muscle strengthening protocols.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-255 mA
Channel Mode . . . . . . . . . . . . . . . . . . . . . . . . . Single, Reciprocal, Co-Contract
Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-1000µsec
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Anti-Fatigue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Off or On
Set Intensity . . . . . . . . . . . . . . . . . . . . . Individual Channel Intensity Setting in
Reciprocal and Co-Contract modes
Cycle Time . . . . . . . . . . .Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, 10/50
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 pps
Ramp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.5 sec, 1 sec, 2 sec, 5 sec
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3 & 4 Option
The Diadynamic waveforms are rectified alternating currents. The
alternating current is modified (rectified) to allow the current to
flow in one direction only.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-80 mA
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
Available on channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1, 2, 3, 4
MF:
(Monophasé Fixe) - Frequency of 50 Hz: phase duration of
10 ms followed by a pause of 10 ms.
DF:
(Diphasé Fixe) - Frequency of 100 Hz: phase duration of 10 ms
followed immediately by another identical phase of 10 ms.
CP:
(Modulé en Courtes Périodes) - 1 second of MF followed
abruptly by 1 second of DF.
LP:
(Modulé en Longues Périodes) - Rhythmical fluctuation
between 2 MF currents.
CP-iso:
(Courtes Periodes Isodynamic) - A combination of MF and
DF waveforms.
CP-id:
Same as CP-iso.
MF+CP:
A period of MF followed by a period of CP.
MF+CP-id:
A period of MF followed by a period of CP-ID.
DF+LP:
A period of DF followed by a period of LP.
DF+CP:
A period of DF followed by a period of CP.
*CC= Constant Current
CV= Constant Voltage
VMS
TM
Diadynamic Waveforms
Supplied by EME Services Ltd