12
Installation
Product Manual - Trifecta
®
XPro Series
Purge and Pre-Charge of the Trifecta XPro Series
Cont.
•
Crack both of the 1/4” Swagelok
®
nuts on the bottom of
the two Wika liquid level gauges (Ll-1) until you can hear
nitrogen gas flowing. Purge through both the high and
low phase connections on both LL gauges for 10 to 15
seconds
•
Ensure that each 1/4” nut is tightened back up after
purging
• Open the purge valves on the safety circuit
• Cycle the diverter valve (DIV-1) to purge out of both
purge valves (V-4) (Do this for tank 1 and tank 2)
• Open both manual vent valves located below the control
box (V-1)
• Purge for one minute
• Close the manual vents (V-1)
•
At any time during the purging process, if the pressure
gets down to 20 PSI, stop purging and open the liquid line
on the bulk tank to let the Trifecta equalize with the bulk
tank again
Note:
Remember to shut the liquid line on the
bulk tank before resuming purging.
• Close the PB Isolation valve on the bottom of tank 1 and
tank 2 (V-3)
• Close VENT/PB Isolation valve on the top of tank 1 and
tank 2 (V-6)
•
Remove 1/2” plug out of the cross on the upstream side
of the 1” P B Magnatrol solenoid valve (SOL-3) on both
tank 1 and tank 2 (See Note 1)
Purge Valves
Diverter Valve
Manual Vent Valves
Plug
Содержание Trifecta XPro Series
Страница 2: ......
Страница 24: ...20 Operation Product Manual Trifecta XPro Series Mode 0 Figure 1 Filling Mode Vent to Storage...
Страница 25: ...21 Operation Product Manual Trifecta XPro Series Mode 0 Figure 2 Filling Mode Vent to Atmosphere...
Страница 26: ...22 Operation Product Manual Trifecta XPro Series Mode 0 Figure 3 Filling Mode Liquid Flow...
Страница 27: ...23 Operation Product Manual Trifecta XPro Series Mode 0 Figure 4 Filling Mode Low Loss...
Страница 28: ...24 Operation Product Manual Trifecta XPro Series Mode 0 Figure 5 Filling Mode Low Loss Vent to Atmosphere...
Страница 29: ...25 Operation Product Manual Trifecta XPro Series Mode Switch 0 to 1 Figure 6 Fill Mode Termination...
Страница 31: ...27 Operation Product Manual Trifecta XPro Series Mode Switch 1 to 2 Figure 8 Secondary Maintains Hold Pressure...
Страница 32: ...28 Operation Product Manual Trifecta XPro Series Figure 9 Primary Pressure Build Mode 2...
Страница 33: ...29 Operation Product Manual Trifecta XPro Series Figure 10 Primary PB Off Mode 2...
Страница 34: ...30 Operation Product Manual Trifecta XPro Series Figure 11 Call for Assist Mode Switch 2 to 3...
Страница 35: ...31 Operation Product Manual Trifecta XPro Series Figure 12 Secondary Assist of Primary Assist Mode 3...
Страница 36: ...32 Operation Product Manual Trifecta XPro Series Mode Switch 3 to 2 Figure 13 Assist Complete Primary Restored...
Страница 40: ...36 Operation Product Manual Trifecta X Series...
Страница 43: ...39 Service Product Manual Trifecta XPro Series...
Страница 47: ...43 Service Product Manual Trifecta X Series...
Страница 48: ...44 Service Product Manual Trifecta X Series...
Страница 53: ...49 Preventive Maintenance Product Manual Trifecta XPro Series Front Isometric view of Trifecta 500 psig...