Specifications:
Overall Dimensions
25 in. (63.5 cm) height x 21 in. (53.3 cm) width x 25 in. (63.5 cm) depth
Weight
140
lbs
(63.6
kg)
Touch Screen
7 in. (800 x 480) TFT-LED glass capacitive touch screen display
One Touch
A combination of vessel recognition and vessel counting sensor technology,
software technology, and applications knowledge that enables a user to
select a One Touch method that matches the sample type. Based on the
sample type, One Touch Technology determines the vessel type and count,
digestion temperature, ramp and hold times, and microwave power input.
PowerMAX
Power control technology provides the optimum amount of energy to the
sample
to
ensure
complete
digestions
Ports
5 USB (1 side, 4 rear), 2 Ethernet, 1 USB-B, 1 RS-232
Sensors
All sensors, including pressure and temperature sensing devices located
within the microwave cavity, are microwave-transparent or shielded to
ensure accurate readings and to eliminate arcing (ignition) hazards.
Languages
Software
available
in
English,
German, French, Italian, Spanish, Chinese and
Japanese
Sample Stirring
In-vessel magnetic stirring of samples
Turntable Design
PerfectCircle™
design
provides
absolute radial symmetry. Turntable operates
in
alternating
or
continuous
mode.
Inlet/Outlet Ports
Standard 0.500” I.D. port or optional 0.3125” I.D. ports for 0.250” (6mm)
tubing
Microwave Cavity
Heavy-duty,
multi-layer
Teflon®
coating
Electrical Requirements
208/230 VAC (207-253 VAC), 60 HZ, 15A @ 208 VAC
220/240 VAC (202-250 VAC) 50 Hz, 15A @ 220 VAC
Magnetron Frequency
2450
MHz
Power Output
1800 watts – continuous power available at all power levels to provide
more control for reactions (IEC 705 Method – 1988)
Magnetron Protection
Solid-state
isolator
(US patent 4,835,354) to protect magnetron from reflected
energy,
ensuring
constant
power
output
Speakers
8
Ω
, 2 W, 86 dB
Printer
Optional on-board thermal printer and USB-B compatible printer port
Safety Features
Three independent door safety interlocks, including an interlock monitoring
system
plus
three
independent
thermal switches, are used in each instrument
to prevent instrument operation and microwave emissions in case of
improper door closure or misalignment. The instrument complies with HHS
standards under 21 CFR, Part 1030.10, Subparts (C)(1), (C)(2) and (C)(3).
Reactiguard continuously monitors the cavity and disables the magnetron if
disturbances occur inside the cavity.
Emissions and Safety Approvals
Conforms to Globally Harmonized EN61010-1 Standard for Safety
Requirements for Electric Equipment for Measurement, Control and
Laboratory
Use
Part
1:
General Requirements (CAN/CSA-C22.2 No.
1010.1-1992).
Patents
CEM microwave systems and vessel designs may be covered by any one of
the
following
US
patents:
04835354, 04080168, 05369034, 04672996,
RE034373, 05230865, 04877624, 04672996, 05206479, 05427741. Other
patents pending.
Teflon is a registered trademark of DuPont.
©2012 CEM Corporation
Содержание MARS 6
Страница 1: ...Operation Manual Microwave Reaction System...
Страница 5: ...iv Ordering Information 92 Specifications 93 Warranty 95...
Страница 13: ...8...
Страница 39: ...34 4 Scroll the Method Information screen to view the remainder of the method information as illustrated below...
Страница 47: ...42 Power Time Graph Temperature Time Graph Temperature in Vessel Graph...