9
14. Select “Line Voltage.”
15. Select and highlight the proper voltage for the instrument – 208V or 230V.
Voltage 200 – 220 Volts – Select 208V
Voltage 221 – 240 Volts – Select 230V
16. Select “OK.”
17. Select the “Save” icon in the upper right corner of the screen to save the proper line voltage.
Note:
The MARS 6 instrument requires 208/230 VAC, 60 Hz, 15A @ 208 VAC or 220/240 VAC, 50 Hz, 15A @ 240 VAC,
single phase power. It requires a non-locking 15 Amp, 250 volt rated receptacle (NEMA 6-15R or a 250 volt. 20 Amp rated
receptacle (NEMA 6-20R).