6
Instrument Setup
1. Carefully open the shipping carton, using caution to avoid puncturing or tearing the foam packaging. Remove the
foam stabilizers from the carton.
NOTE
Retain all packing material for use if returning the instrument to the manufacturer for service.
2. Remove the outer box.
3. With at least two people for lifting, locate the hand-hold cutouts on each side of the inner box and lift the
instrument from the shipping carton and place it on the floor in front of the laboratory bench where the instrument
is to be placed.
4. Open the inner box and remove all packaging materials including the plastic wrap from the instrument.
5. Lift the instrument onto the laboratory bench within 8’ of the exhaust source.
Note:
The instrument requires 26” clearance.
6. Remove the contents from the instrument cavity. Install the elbow and exhaust hose on the back of the
instrument. Plug the power cord into the MARS 6 and the dedicated electrical outlet.
Verify that all accessories listed below have been included:
Detachable Power Cord
Exhaust Hose Kit
8’ x 3” Hose
3”
Elbow
Hose
Clamp
Drain Line Fitting
5’ Silicone Drain Tubing
3 Plastic Pipe Plugs
15A Fuse (2)
Operation Manual on USB flash drive
Quick Start Guide
General Guidelines to Assist with Microwave Digestion Method Development
Power Cord (60 Hz instrument)
Note:
Optional items such as digestion vessels, turntable, or capping system may have been included in the shipment.
Carefully check the packing list(s) and the contents of shipping cartons to verify that all items listed are included. If any
items are missing, notify CEM Corporation or a local subsidiary or distributor in the appropriate area of any discrepancies.
7. Position the power switch to the “On” position.
8. Once the instrument initializes, the following screen will appear.
9. Select
“OK.”
Содержание MARS 6
Страница 1: ...Operation Manual Microwave Reaction System...
Страница 5: ...iv Ordering Information 92 Specifications 93 Warranty 95...
Страница 13: ...8...
Страница 39: ...34 4 Scroll the Method Information screen to view the remainder of the method information as illustrated below...
Страница 47: ...42 Power Time Graph Temperature Time Graph Temperature in Vessel Graph...