   Operator’s manual 

 

 

October 2012

                                                                                                                                 

16 

 

The information about the condition of the corresponding cell is shown when one of 
the numerical keys <1>, <2> and <3> is pressed. If the chosen cell is not used for 
any tests, the available cells are shown. (an asterisk positioned on the right of the 
name is used to indicate the unavailability). If there is an analysis on-going on the 
cell, there is an indication of the type of analysis in progress and the messages for 
the operator are displayed in order to proceed with it. 

 

The status of the cells is shown by pressing the numerical <0> key. Pressing it 
again, all tests are shown. 

 

In the second modality the list of all the available analysis is displayed for the 
instrument (an asterisk positioned on the right of the name highlights the 
unavailability). Pressing the numerical <0> key the visualisation of the condition of 
the cells is effectuated. 
 

4.4 Test selection and execution 

On 

stand-by dialog

 press the <0> key to display the list of tests available. Position 

the cursor with arrow key up/down to select the test to run. Press enter key to 
confirm, the first test dialogue is shown on display. 

 

It is also possible to run a test on a specific reading cell. E.g. press 1 for reading cell 
1 and all tests that could run are shown. Select the test you want and press <enter> 
to confirm. The first test dialogue is shown on display then. 

 

Regarding to the chosen tests, the instrument requests the intervention of the 
operator to insert the test tubes and to start the reading. The reading cell that must 
be used is indicated by a green LED. During reading, the green LED changes to red. 

 

 

In all cases to proceed a test, go as described in the detailed Test 
Application Instructions.   

 

 
Results are shown automatically on display and printed out. After that you have the 
possibility of reprinting them or press <down arrow> key to turn to the list of test.  
 

Cancellation of a test

: press <down arrow> key to cancel a running test. 

 

Handling of reagents 

Read carefully the information provided in the 

Test Application Instructions! 

 

 

4.5 Printing of the results 

The results of the analysis are shown on the display and printed out on paper. 
The print-out format as an example: 
 

Содержание FOODLABFAT

Страница 1: ...Operator s manual October 2012 1 Instrument for the Quality Control of Oils and Fats In Vitro Analysis Made in Italy CDR s r l Firenze Italy Development Manufacturing Sales Technical Service...

Страница 2: ...arch 2011 J Schenne 2nd edition R3 October 2012 J Schenne 3th edition USA Document made by Ing Buero fuer Bio Prozessanalytik J Schenne Ohlstuecken 9 21033 Hamburg Germany info foodlab arcor de In cas...

Страница 3: ...and 6 1 3 2 Safety of the product 7 2 System description 9 2 1 The FOODLABfat analyser 9 2 2 Instrument nameplate 9 2 3 Operating parts 10 2 4 1 Key folio layout and display 10 2 3 2 Reading cells 10...

Страница 4: ...f the results 16 4 6 Standardisation of tests 17 4 7 Menu description 17 4 7 1 Edit dialogue 17 4 7 2 Setup dialogue 18 4 7 3 Auto start dialogue 19 4 7 4 Printer dialogue 19 4 7 5 Info dialogue 20 5...

Страница 5: ...way and using the correct methods as agreed by the manufacturer If the operator finds a disagreement between anything described in this manual and the instrument he must advise the manufacturer immed...

Страница 6: ...rixes different from those indicated in the methodologies The instrument is specifically calibrated during construction to test the biological matrixes indicated in the methodologies The instrument ca...

Страница 7: ...2 Safety of the product 1 3 2 1 European Directives consulted during the design stage The instrument has been designed and constructed following the health and safety requirements of the directive Tab...

Страница 8: ...requisites of the norms EN 50081 1 Electromagnetic compatibility emission residential environment light industry EN 55022 A Radio disturbance measures on equipment for computer technology EN 50082 2...

Страница 9: ...Reading cells Figure 2 1 Modules 2 2 Instrument nameplate The instrument has a nameplate that displays the CE mark and identification data Figure 2 2 Nameplate Important Never remove the nameplate fr...

Страница 10: ...Up Move the cursor upwards on the display stop number of samples Down Move the cursor downwards on the display interruption of a running test Power ON OFF the instrument Numerical keys 0 9 Enter Confi...

Страница 11: ...ep it clean don t touch it with fingers Figure 2 5 CDR test cuvette 2 3 4 CDR Reagent kit Test kit contains 10 pre filled cuvettes in an aluminium bag For example Test name Acidity Code 300125 Lot rea...

Страница 12: ...in its carton must be stored in a closed environment within the temperature range 20 C to 70 C 4 F to 158 F and a relative non condensed humidity of 20 to 90 The storage place must not be exposed to c...

Страница 13: ...ed as follows A Take the FOODLABfat out of the carton and put it on a plane surface in such a position that it is not exposed by mechanical vibrations or extreme temperature fluctuations or excessive...

Страница 14: ...peration procedure The FOODLABfat is an instrument equipped with dedicated software for the deter mination of parameters on different oils and fat extracts and alimentary matrixes of various natures T...

Страница 15: ...the enter key to proceed the calibration process After that the results are shown on display All installed optics must be ok In case of differences refer to chapter 5 Error messages Press enter key t...

Страница 16: ...ion the cursor with arrow key up down to select the test to run Press enter key to confirm the first test dialogue is shown on display It is also possible to run a test on a specific reading cell E g...

Страница 17: ...position ANL STD to STD ANL then you will be requested to define up to 14 points After measuring results are shown on display together with K and Q values and the regression coefficient R These values...

Страница 18: ...nal value SMP Timer definition of the incubation time up to 15 minutes for an activation prior reading of the sample To change this time press the enter key and successively the up arrow key and the d...

Страница 19: ...7 3 Auto start dialogue In this mode it is possible to insert the time of automatic switching on of the FOODLABfat instrument if it is connected to the mains and is ready for testing after 45 minutes...

Страница 20: ...4 7 5 Info dialogue After selection the following information is shown Registration represents the registration of the instrument Firmware represents the version of programme installed Operator operat...

Страница 21: ...own and then switch it on again If the problem persist please contact the technical assistance department for repairing In the following are summarized the instructions to be followed in case specific...

Страница 22: ...2 Rif key has been selected 6 Maintenance 6 1 Recommended actions and precautions The FOODLABfat instrument is easy to use and it is normally maintenance free Do not spill any liquid over the instrum...

Страница 23: ...your distributor 7 2 Spare parts CDR AEP 119 1 x roll of thermal printer paper 57 mm diam 45 mm CDR ALP 004 external power adapter 100 230V AC to 12V dc CDR 225 058 protective cover for reading cells...

Страница 24: ...Operator s manual October 2012 24 7 3 Declaration of conformity...

Страница 25: ...d 4 x 3 positions Cuvettes Volume at triangular mark 1000 l mixed polymer for optical use Cups Polystyrol made Sample volume depending on test Throughput depending on test Processor MC 68331 micro pro...

Страница 26: ...READ_CHECK chk cc check reading cells cc OK KO DATE date in the days month years format TIME time in the hours minutes format CALIBRATION calibration DATE date in the days months years format TIME tim...

Страница 27: ...he rear enable to connect the analyser to a computer FOODLABfat can be connected to LIMS Laboratory Information Management System and the XML Extensible Markup Language protocol 7 6 Waste disposal The...

Страница 28: ...tes of particular interest protected by the norms in force ATTENTION Do not hesitate to contact your distributor about any question that is inherent to the demolition procedures of the instrument and...

Страница 29: ...Operator s manual October 2012 29...

Отзывы: