This document is the property of Tecora and can not be duplicated without permission
Ref : EN MU DSM 46 V2
Page 14 of 49
3.1.3 Volumetric Sampling, set the sampling Start and volume to realize
•
Move the cursor using the arrows and select
“
Volumetric Sampling”
and then press the Enter
key.
The sampling program mask is displayed
•
Date and Hour
are the effective sampling start date
and hour. When you enter this screen, the date and
are the current ones.
In case you want to start immediately, you can skip
those settings.
•
Volume
is the volume in liters to realize during the
sampling.
The minimum sampling time should be not less than 5 minutes while the maximum is 99
hours and 59 minutes.
•
Norm. Temp
is the used temperature, expressed in °C, to normalize the volumes. The
default value is 25°C.
•
Press. Atm
. is atmospheric pressure, expressed in kPa, used from the instrument. The
default value is 101.3 kPa.
Note: Atm Press. Isn’t the normalizing pressure (which is always defined as 101.3 kPa) but the
atmospheric pressure at which the instrument operates .
To start the samplings, move the cursor on START and press the Enter key.
More information about the sequential and intermittent sampling see the
Chapter 6 –
Sequential Sampling -
SAMPLE TIMING
start
start
Date 15-05-03
hour 08:35
Volume
liters
150.00
Norm. Temp. °C
25.000
Atm Press. kPa
100.2
START
SEQUENTIAL ESC
MENU PRINCIPALE
Timed Sampling
Volumetric Sampling
Test Pump OFF
Utility Menu
Explore Memory
ESC