background image

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

    
    
 

 

 

 
 

Bravo Pump 

Plus / Basic 

 

USER'S MANUAL 

MANUAL 

Содержание tecora BRAVO basic

Страница 1: ...Bravo Pump Plus Basic USER S MANUAL MANUAL ...

Страница 2: ...rsion 2a Pages 2 of 49 Date 2019 03 18 Date Written by Date Validated by 2019 03 18 VLC 2019 03 18 VLC 2019 03 18 2a All Modification into CDL 2017 04 24 2 All New presentation add timer GRASSLIN 2006 04 01 1 3 All DATE Version PAGE Modifications ...

Страница 3: ...12 3 1 2 Timed Sampling set the sampling Start and Duration 13 3 1 3 Volumetric Sampling set the sampling Start and volume to realize 14 3 2 Data displayed during the sampling 15 3 3 Pause during the sampling 16 3 3 1 Timed Sampling 16 3 3 2 Volumetric sampling 16 3 4 Sampling Final Report 17 4 Utility Menu 18 4 1 Test EV 18 4 2 Set datary 18 4 3 Set the Baud Rate 19 4 4 Download the memory using ...

Страница 4: ... 27 8 1 SMS Commands 27 8 1 1 Start the sampling 27 8 1 2 Sampler actual status 28 8 2 Anomalies in the SMS transmission 28 9 BRAVO Basic 29 9 1 Instrument Description 29 9 2 Operation Instructions 30 9 2 1 Power 30 9 2 2 Power On 30 9 2 3 Flowrate Setting 30 9 3 Timer Programming 31 9 3 1 Timer Description 31 9 3 2 First Program 31 9 3 3 Normal Operation 32 9 3 5 Programming the device 32 ...

Страница 5: ...mmediately Test The instruments are tested in our laboratories at Villebon sur Yvette ZI La Prairie and a test certificate is released Warranty The warranty cover the instruments for 12 months from the test passed date The warranty doesn t cover the following damages to the instruments occurred during the transport damages to the instruments due to negligence or wrong use wearable parts batteries ...

Страница 6: ...ons a good power supply with the ground and no disturbs is needed In case you use an extension cable for the power supply it s necessary that you connect only the ECHO PM to this power supply cable On the power supply line of the ECHO PM there is an anti disturb filter The socket type is a two phases plus the ground with protection fuse on each phase and a bipolar ON OFF switch with ON LED Power S...

Страница 7: ...y data and logger Sequential Interface Optional Piston Pump Range 1 60 l min New Controller for samplings program display data and logger Sequential Interface Optional BRAVO M Basic BRAVO R Basic BRAVO H Basic Diaphragm Pump Range 0 15 35 l min weekly based programmer Digital Thermometer 3 m3 h Rotary vane Pump Range 0 2 35 l min weekly based programmer Digital Thermometer 6 m3 h Rotary vane Pump ...

Страница 8: ...t permission Ref EN MU DSM 46 V2 Page 8 of 49 BRAVO Curve caratteristiche Performance curves BRAVO M BRAVO R 0 10 20 30 40 50 60 70 80 90 100 0 10 20 30 40 50 60 70 Portata Flow qV l min Vuoto Vacuum KPa BRAVO H BRAVO M BRAVO R Bravo curve characteristics Performance curves ...

Страница 9: ...cription 1 Membrane keypad and back lighted display 2 Cover with key lock 3 Inlet quick connector 4 Vacuum meter in line with the sampler gas 5 Dry Gas Meter 6 Flow rate regulators 7 RS 232 Interface 8 Flow Meters 9 Sequential interface connector Optional 10 Power supply socket with lighted switch and protection fuse 1 2 7 8 9 5 3 4 10 6 ...

Страница 10: ...e duplicated without permission Ref EN MU DSM 46 V2 Page 10 of 49 2 1 1 Keyboard and Display Keyboard Pause key Move the cursor to the left and move on upside raw Move the cursor to the right and skip a raw Skip a raw Accept the selection Numeric keyboard ...

Страница 11: ...Tecora or your local dealer The standard supply includes 1 BRAVO Plus 2 Power cord 3 3 mt Rauclair tube 4 Quick connector 5 2 spare fuses 6 Calibration certificate 7 User Manual 2 3 Accessories The Instrument can be equipped with the following accessories and spare parts Accessories and Spare Parts P N Sequential Sampling Interface AA99 000 9903SP Serial Cable 999NI012 Software data acquisition Do...

Страница 12: ... 3 1 1 Flowrate Setting Before to program the sampling set the flowrate to the desired value Move the cursor using the arrows and select Test Pump OFF and press Enter The pump will turn on and will be possible to set the flowrate using the flow regulators pos 6 and reading the value on the flow meters placed on the front of the instrument Once set the flowrate go back with th cursor on Test Pump O...

Страница 13: ...s while the maximum is 99 hours and 59 minutes Norm Temp is the used temperature expressed in C to normalize the volumes The default value is 25 C Press Atm is atmospheric pressure expressed in kPa used from the instrument The default value is 101 3 kPa Note Atm Press Isn t the normalizing pressure which is always defined as 101 3 kPa but the atmospheric pressure at which the instrument operates T...

Страница 14: ...s than 5 minutes while the maximum is 99 hours and 59 minutes Norm Temp is the used temperature expressed in C to normalize the volumes The default value is 25 C Press Atm is atmospheric pressure expressed in kPa used from the instrument The default value is 101 3 kPa Note Atm Press Isn t the normalizing pressure which is always defined as 101 3 kPa but the atmospheric pressure at which the instru...

Страница 15: ...ffective sampling time pump ON from the start Volume C V is the volume expressed in m3 sampled until this moment and referred to the dry gas meter conditions Volume std is the volume expressed in m3 sampled until this moment and referred to the normalized conditions temperature and pressure where the temperature is the one selected during the sampling program and the pressure is 101 3 kPa Temp C V...

Страница 16: ...ampling In this menu it s possible to modify the sampling Duration and the Cycle for sequential samplings Selecting STOP SAMPLING the operation will be interrupted manually 3 3 2 Volumetric sampling In this menu it s possible to modify the Volume and the Cycle for sequential samplings Selecting STOP SAMPLING the operation will be interrupted manually ATTENTION The instrument manages the power loss...

Страница 17: ... ones related to the last sampling Pressing the arrow keys the other memorized reports are displayed Pressing the Enter key you will go back to the main menu Displayed data Sample 1 of 1 identify the sequential samplings Data Str is the sampling starting date time Elapsed time is the effective sampling time Volume c v is the total volume measured by the dry gas meter Volume Std is the volume repor...

Страница 18: ...l ways open and close will be done 4 2 Set datary The instrument is equipped with a permanent clock datary with a backup battery To modify the settings select Set Datary from the Utility Menu and press Enter The following display will appear Use the arrow key to select the desired field Use the keyboard to edit the date and clock To memorize the new date and exit select UPDATE and press Enter UTIL...

Страница 19: ...nd managing memory data it s supplied with a serial cable the P N is AA99 099 9900SP 4 4 1 Measure final report Selecting RS232 Mem in the Utility menu the logged data are sent to the RS 232 port The format is the following Date Start hour Date Stop hour Cycle Duration Sample n of n Effective sampling time ET hhh mm ss A string followed by with the following data Programmed Ambient Pressure Pam kP...

Страница 20: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 20 of 49 Power return gg mm aa hh mm ss ...

Страница 21: ...rature sensor readings Tc v Dry Gas Meter Temperature Pressing the 1 key it s possible to enter the temperature value measured by the reference instrument Pressing ENTER you ll go back to the previous menu The calibration of those parameters requires specific equipment temperature calibrators and tools Attention if the calibrator reported value differs a lot more than 20 from the instrument value ...

Страница 22: ...ase a wrong calibration value is entered or before to proceed with a new calibration it s possible to load the default calibration values Selecting Load Default Calibration the default factory values are automatically restored UTILITY MENU Test EV OFF Set datary Baud 9600 Download Memory Delete Memory Measures Calibrations Load Default Calibration ESC ...

Страница 23: ...ling and the next called Cycle The first part of the program is similar to the one already discussed in the charter 3 2 Program a sampling In this example we want to program a daily sampling of 24 hours to be executed on alternate days Once edited the data press SEQUENTIAL to enter in the sequential program section Sample n is the sequential sampler programmed 2 Cycle is the interval between the s...

Страница 24: ...ty of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 24 of 49 The Cycle value must be necessarily equal or greater than the Duration Duration 24h Duration 24h Cycle 48 h Cycle 48 h 1 sample 2 sample ...

Страница 25: ... the campaign over longer times In this case the intermittent option allows extending the sampling period over more days limiting the effective sampling time to the effective sampling media capacity Practically the Duration of the sample can be set to 96h 4 days but the effective sampling time is limited to 24 hours To do this Set the duration to 96 hours and then select the sequential option The ...

Страница 26: ...s messages given by the instrument Available memory 20 The available memory space is low Delete the memory Change the Battery Indicates that the backup battery has to be charged Leave the instrument under charge 7 Measuring sensors Characteristics Temperature Sensors Pt 100 Thermo resistors Range Norm DIN IEC 751 Class B 1983 Range 30 500 C Precision 0 5 C 30 50 C 0 8 C 50 100 C 1 2 C 100 C ...

Страница 27: ...essages It s possible that messages with the correct syntax but containing wrong characters or numbers could create operational anomalies in the instrument 8 1 1 Start the sampling The start sampling command launches immediately the sampling To use this command the instrument must be in the stand by condition no sampling The instrument will start the sampling immediately it s not possible to set a...

Страница 28: ...received command STATUS command sent to the sampler SAMPLING ON status of the instrument 8 2 Anomalies in the SMS transmission Messages arriving from the phone card provider or from other portable phones could cause the block of the message receiving on the sampler This possibility could lead to a block if the frequency of those messages is very high and the instrument is not able to delete them I...

Страница 29: ...9 9 BRAVO Basic 9 1 Instrument Description 1 Timer Keyboard and LCD Display 2 Cover with key lock 3 Sample Inlet with quick connector 4 Vacuum meter 5 Dry Gas Meter counter 6 Flow regulator knobs 7 Temperature indicator 8 Rotameters 9 Sampled gas outlet 10 Power supply switch with protection fuse 1 2 7 8 9 5 3 4 10 6 ...

Страница 30: ...nnector to the sampler pos 3 press the ON OFF key on the timer for further information about the timer programming see paragraph 9 3 to start the pump and proceed with the flowrate regulation 9 2 3 Flowrate Setting There are two flowrate regulators on the sampler one is high flowrates dust asbestos lead samplings and one for low flowrates gas solvents etc with the exception of the BRAVO H Basic wh...

Страница 31: ...ing proceed with the instrument cold reset General notes about the programming To quickly exit from the program in any moment just press the key If any key is pressed for 30 seconds automatically the timer will return to the normal operation Reset Partial Reset with this function the date is set to the default 2000 01 01 the hour is set to the default 00 00 the RANDOM program if active is set to O...

Страница 32: ...nfirm the selection and skip to the next field in cycle MINUTES HOUR YEAR MONTH DAY SECONDS Note The indication for the day of the week zone 4 will be updated automatically 4 To memorize the values and go back to the normal operation press again 9 3 5 Programming the device The timer can be programmed for Turn on ON and Turn Off OFF events The maximum events number is 25 1 Press the PRG to enter i...

Страница 33: ...m Press OK to skip the days selection and go directly to the day number selection The parameter under selection is indicated by blinking dashes since it s not needed to specify the month and year press OK twice and go to the day number setting To set the day use the key and to confirm the selection press OK Press again OK to go to another event program Press to return to the normal operation Yearl...

Страница 34: ... event it s not possible to change a turn on event in turn off Deleting a program Press PRG to access to the program mode Select the program to be deleted using the OK key To permanently delete the program press and keep pressed the PRG key for about 5 seconds to confirm the successful deletion the message CLEAR will be displayed Once deleted a program the next program if present or a new one is d...

Страница 35: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 35 of 49 Timer programming type GRASSLIN ...

Страница 36: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 36 of 49 ...

Страница 37: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 37 of 49 ...

Страница 38: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 38 of 49 ...

Страница 39: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 39 of 49 ...

Страница 40: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 40 of 49 ...

Страница 41: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 41 of 49 ...

Страница 42: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 42 of 49 ...

Страница 43: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 43 of 49 ...

Страница 44: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 44 of 49 ...

Страница 45: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 45 of 49 ...

Страница 46: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 46 of 49 ...

Страница 47: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 47 of 49 ...

Страница 48: ...This document is the property of Tecora and can not be duplicated without permission Ref EN MU DSM 46 V2 Page 48 of 49 ...

Страница 49: ...03 66 69 Fax 33 0 4 91 46 58 47 S A S AU CAPITAL DE 24 000 R C S Marseille B377847546 SIRET 37784754600027 TVA FR 68377847546 Atelier CDL Tecora 10 rue de la Prairie 91140 Villebon sur Yvette COMBUSTION ÉMISSION DEPARTMENT INDUSTRIAL CONTROL DEPARTMENT HYGIENE HEALTH AND ENVIRONMENT DEPARTMENT SERVICE AND LOGISTIC DEPARTMENT ...

Отзывы: