TROUBLESHOOTING GUIDE
WARNING: Before calling the Service check the following guide and the enclosed
alarms table.
FAULT
CAUSE
REMEDY
Plug in the socket, green
power switch in
I
position but
everything is Off (including
the green light on the switch)
a)
No power supply
b)
Burned fuses
a
) Check the power supply
b
) Replace the fuses in the fuse box
within the mains socket
Remove the power cable from the
socket before replacing the fuses
The power cable must be removed from the socket before replacing the fuses.
If when switching on or during the first cycle the fuses should burn straight away, call the
Service in order to remove the fault which causes the fuses to burn.
MAX. THERMOSTAT - (42) e (43)
The sterilizer is fitted with 2 thermostats which trip in case of
overheating. To reset them working leave the door open and
wait a few minutes for the temperature to drop, then remove
the cap if there is one and press the thermostats’ reset button
located at the back of the sterilizer.