C.B.M. S.r.l. a socio unico - Medical Equipment –
Via Castello 10, 26038 Torre dé Picenardi (CR) - Italy
Rev. 06 – 18/02/2019
Manuale SST1700-2200 r06 CE-1282 EN
page 17 of 34
♦
WATER LOADING SELECTION
- Use of the “WATER LOADING SELECTION"
function
It allows to select the water loading mode.
To modify the load kind a Password shall be entered.
The pre-set password is “
00000
”, it can be modified and a new Password can be saved
through the PASSWORD menu.
-
MANUAL:
(front side) when water loading through the front side of the appliance (pre-
set) is planned, through cans and / or vessels
-
AUTOMATIC:
(rear side) when you wish to load water by connecting the appliance to
the rear fitting (50) and to the proper demineralizers producing demineralized water by
using the main network water.
♦
LANGUAGE
- Use of the “LANGUAGE" function
By selecting this key the favourite language can be set chosen among the suggested ones:
ITALIAN, ENGLISH, FRENCH, GERMAN, SPANISH.
To change the language a Password shall be entered.
The pre-set password is “00000”, it can be modified and a new Password can be saved
through the PASSWORD menu.
♦
Enable / disable USB Flash Drive
- Use of the “USB – enable/ disable" function
By selecting this key the use of the USB flash drive can be enabled or disabled.
The USB flash drive is pre-set as Enabled.
This function is usually used by the responsible person when several persons are using the
appliance.
To enable or disable the USB flash drive a Password shall be entered.
The pre-set password is “
00000
”, it can be modified and a new password can be saved
through the PASSWORD menu.
♦
SERVICE
- Use of the “SERVICE" function
Within this menu several setting and test functions are present to be used only by an
authorized technician.
The authorized technician can get to the SERVICE menu which is protected by a proper
“service password”.
Among the operations it is possible to increase or reduce the parameters of conductivity
acceptance in µS (Micro Siemens) of the demineralized water to be used, pre-set as < 80
µS. The increase of this parameter cancels the warranty.