C.B.M. S.r.l. a socio unico - Medical Equipment –
Via Castello 10, 26038 Torre dé Picenardi (CR) - Italy
Rev. 06 – 18/02/2019
Manuale SST1700-2200 r06 CE-1282 EN
page 12 of 34
GENERAL INSTRUCTIONS
(13) – GREEN POWER SWITCH:
Located under the display cover; when pressed switches the appliance ON enabling its use.
(28) – OPEN DOOR
To open the door, press the key on the display and pull the door slightly.
(28) – CLOSE DOOR
Manually push the door till it hits the sterilizing chamber, then press the key on the
display; after a few seconds the DOOR CLOSED
icon will appear on the display.
The door will be fully locked only after having pressed the START key.
(10) - (12) WATER TANKS FILLING/EMPTYING pipe-fittings:
Opening the cover below the display reveals 2 pipe-fittings located in the lower area and
connected to the two tanks of the appliance.
♦
Coupling with Blue or White left button: to be used to fill the Clean Water tank.
♦
Coupling with Grey or Black right button: to be used to empty the Waste Water tank.
(12) – FILL CLEAN WATER – Blue or white button
coupling
By means of the tank filling function it is possible to
refill automatically from a
clean water can.
Fit the appropriate hosepipe (supplied) by pressing the pipe-fitting onto the blue “
Fill Clean
Water Tank”
coupling.
Place the other end of the hosepipe in the demineralized water can making sure it touches
the bottom.
Press the Right direction key on the display then press the Fill WATER key.
The tank holds a maximum of 3 litres of water; when the tank is full, FILLING will stop
automatically and the “Tank Full” icon will appear on the display.
When the water in the can ends before the tank is completely full, please press the
key on the display to interrupt water loading and to avoid overheating and/or breaking of the
loading pump