18

Routine maintenance entrusted to specially instructed 
surgery staff 

(draw. 3 page 43)

•  Electrical shock risk : also 230 V 

 can be lethal.  

•  General danger sign.  

•  Biological danger, risk of infections from epidemic diseases.  

•  Compulsory direction of fl ow or of rotation.  

1) 

After each surgical operation or particularly long operations, aspirate abundant clean 
water to rinse the suction piping.  

2)

 At the end of each working day, clean and disinfect the appliance using a specifi c 

product for dental aspirators and amalgam separators. We suggest, for example, Puli-Jet 
plus (for dilution with water, refer to our directions). Do not use aggressive or foaming 
products.

3)

 The dirty fi lter (17) must be replaced every day. Handle it with care to prevent contamina-

tion and the return of solid trapped particles into the fi lter. Any material trapped in the 
fi lter must be collected and disposed of according to the regulations in force.

4)

 We recommend placing one disinfectant antifoam tablet in the dental unit secretion fi lter 

every morning before starting the aspiration. 

5)

 In case the appliance is treated with products which are different from the ones recom-

mended by the importer, the correct working of the equipment might be compromised 
and faults could occur. 

6)

 Check that there are no liquid leaks. In case of working anomalies or alarms, contact 

the technician. 

Faults report and possibile solutions

The amalgam container fi lling levels are the only alarms directly linked to the Hydro-
cyclone.
Below is information useful to inform the reader about possible problems that may 
cause the aspiration to stop and that may compromise the working of the “Separatore 
Centrifugo Compatto” for the Hydrocyclone.

With the circuit board AC58 (draw. 6 page 46)

1)

 If led number 32 is off it means that there is no power supply: check that the machine 

switch and the master switch are on, in case they are on but light 32 is still off, contact 
the technician.

2)

 The sound of the alarm and the alternation of the fl ashing of the yellow led 18 (double 

fl ashing) and the one of the red led 20 (one single fl ash) indicate that the AC58 ampero-
metric control system has detected an anomaly in the centrifugal separator current 
absorption. Such an anomaly could have different causes: mechanical blockage or 
a drain obstruction (8) (draw. 3 page 43) of the Hydrocyclone or a temporary voltage 
decrease. 

  Disconnect the appliance from the mains for a few seconds to reset the alarms and to 

start the appliance again. In case the alarm appears again, contact the technician. 

With the circuit board AC80 (draw. 6 page 46)

1) 

If the display (22) is off it means that there is no power supply: check that the machine 
switch and the master switch are on, in case they are on but the machine is still off, 

Содержание IDROCICLONE ISO 18

Страница 1: ...334 DE 0351372 ES 0351372 FR 0351372 GB 0351372 EP 0791751 US 5 779 443 CH 0791751 DE 0791751 ES 0791751 FR 0791751 GB 0791751 PT 0791751 AU 93321 ES 107358 FR 222 394 395 PENDING PATENT CATTANI IT M...

Страница 2: ...SE 15 INSTALLATION 16 MOUNTING 17 NOTICES 17 GENERAL WARNINGS AND BIOLOGICAL RISK 17 ROUTINE MAINTENANCE 18 FAULTS REPORT AND POSSIBILE SOLUTIONS 18 EXTRAORDINARY MAINTENANCE AND ROUTINE ANNUAL FIVE Y...

Страница 3: ...12...

Страница 4: ...or their functioning must never be modified Field of use for the German market only According to the annex no 50 of the General Framework of the Administrative Standard regarding the minimum requireme...

Страница 5: ...will find some details listed below SEPARATORE CENTRIFUGO COMPATTO Name SEPARATORE CENTRIFUGO COMPATTO Electric power supply 230 240 V 1 50 Hz 2 8 A 230 400 V 3 50 Hz 2 2 1 3 A Maximum allowed negativ...

Страница 6: ...won t be possible to start the machine again unless the amalgam collection container 5 is replaced Every time an alarm causes the machine to stop a volt free clean contact closes terminals 25 26 draw...

Страница 7: ...since the contained material is toxic and contaminating for epidemic diseases we recommend to handle it with care To receive a list of authorized companies with licence for transport and disposal of t...

Страница 8: ...e installer should train the surgery staff with practical demonstra tions for amalgam container replacement and filter cleaning on new and therefore not contaminated machines Installation connection m...

Страница 9: ...liquid leaks In case of working anomalies or alarms contact the technician Faults report and possibile solutions The amalgam container filling levels are the only alarms directly linked to the Hydro...

Страница 10: ...15 inserting a metallic piece between the two photodiodes The Hydrocyclone control panel or the display will show a visual alarm to indicate that the amalgam level has reached 95 or 100 3 Remove and...

Страница 11: ...tant Remove the liquids and residual debris which are inside the appliance place it in a poly ethylene bag seal and pack it in packaging constructed of 3 layer corrugated board Important notices The m...

Страница 12: ...l ssigkeit Ricircolo Re circulation Umlauf Tubo in ingresso separatore Separator hose inlet Abscheidervorgang Uscita liquido depurato Processed liquid outlet Ablauf der gereinigten Fl ssigkeit Illustr...

Страница 13: ...Fig 2 Draw 2 Abb 2 SEPARATORE D AMALGAMA IDROCICLONE ISO 18 MONTAGGIO AMALGAM SEPARATOR HYDROCYCLONE ISO 18 MOUNTING AMALGAMABSCHEIDER HYDROZYKLON ISO 18 MONTAGE 42...

Страница 14: ...MOUNTING AMALGAMABSCHEIDER HYDROZYKLON ISO 18 MONTAGE all aspiratore to the aspirator zur Saugmaschine D 50 mm ai riuniti to the dental units zu den Behandlungseinheiten D 50 mm Separatore aria acqua...

Страница 15: ...Fig 4 Draw 4 Abb 4 SEPARATORE D AMALGAMA IDROCICLONE ISO 18 MONTAGGIO AMALGAM SEPARATOR HYDROCYCLONE ISO 18 MOUNTING AMALGAMABSCHEIDER HYDROZYKLON ISO 18 MONTAGE 44...

Страница 16: ...ISO 18 COLLEGAMENTO AMALGAM SEPARATOR HYDROCYCLONE ISO 18 CONNECTION AMALGAMABSCHEIDER HYDROZYKLON ISO 18 ANSCHLUSS all aspiratore to the aspirator zur Saugmaschine all aspiratore to the aspirator zu...

Страница 17: ...ndicazione d allarme come visualizzata sul display del circuito AC80 Alarm indication as shown on the display of the circuit AC80 Alarmanzeige wie auf dem Display der Platine AC80 angezeigt Report Cod...

Страница 18: ...SEPARATORE D AMALGAMA IDROCICLONE ISO 18 ISTRUZIONI PER L USO AMALGAM SEPARATOR HYDROCYCLONE ISO 18 DIRECTIONS FOR USE AMALGAMABSCHEIDER HYDROZYKLON ISO 18 GEBRAUCHSANWEISUNG Fig 7 Draw 7 Abb 7 47...

Страница 19: ...ROCICLONE ISO 18 MANUTENZIONE AMALGAM SEPARATOR HYDROCYCLONE ISO 18 MAINTENANCE AMALGAMABSCHEIDER HYDROZYKLON ISO 18 WARTUNG Fig 8 Draw 8 Abb 8 Corpo metallico per test Metallic piece for checking Met...

Страница 20: ...SEPARATORE D AMALGAMA IDROCICLONE ISO 18 MANUTENZIONE AMALGAM SEPARATOR HYDROCYCLONE ISO 18 MAINTENANCE AMALGAMABSCHEIDER HYDROZYKLON ISO 18 WARTUNG Fig 9 Draw 9 Abb 9 49...

Страница 21: ......

Страница 22: ...334 DE 0351372 ES 0351372 FR 0351372 GB 0351372 EP 0791751 US 5 779 443 CH 0791751 DE 0791751 ES 0791751 FR 0791751 GB 0791751 PT 0791751 AU 93321 ES 107358 FR 222 394 395 PENDING PATENT CATTANI IT M...

Отзывы: