Mounting 

(draw. 5 page 45)

•  Electrical shock risk : also 230 V 

 can be lethal.  

•  General danger sign.  

•  Biological danger, risk of infections from epidemic diseases.  

•  Compulsory direction of fl ow or of rotation.  

Distances

The maximum distance does not depend on the Hydrocyclone but on the suction pump, 
we therefore recommend that you refer to the instructions of the suction pump supplied 
by the manufacturer. 

Positioning

The Hydrocyclone can be installed in the following layouts:

1)

 installation on the same level of the dental units

2)

 installation on the lower level

The Hydrocyclone must be easily accessible from all sides. Leave a free space on the 
top and on the front of the unit to facilitate maintenance procedures.
Liquids drain due to fall, therefore we recommend to provide a fall towards the sewage 
to facilitate the draining. It is possible to place a small trap, provided that it does not 
obstruct the draining. 

Notices

After installation, the installer should train the surgery staff with practical demonstra-
tions for amalgam container replacement and fi lter cleaning on new and therefore 
not contaminated machines. 
Installation, connection, mounting, maintenance and other work on the amalgam 
separator must be entrusted to technicians who have been specially trained in 
the dental fi eld and who are authorized by the manufacturer. The aspirated liquid 
is contaminating and infected, therefore we recommend to take all the necessary 
precautions to prevent personal and environmental contamination. Bad running 
can also lead to contamination. 

General warnings and biological risk

•  Before starting any servicing operation on appliances which have been used, carry 

out a series of rinses with a suitable disinfectant.

• Disconnect the centrifugal separator from the mains supply and padlock the 

isolating switch, if present.

•  Before starting any maintenance work, wear gloves, goggles, mask and disposable 

overalls. 

The centrifugal separator pressurises aspirated liquid and forces it inside the 
Hydrocyclone.
In case of a failure of a pipe or of a component under pressure biological risk is 
more serious.

Содержание IDROCICLONE ISO 18

Страница 1: ...334 DE 0351372 ES 0351372 FR 0351372 GB 0351372 EP 0791751 US 5 779 443 CH 0791751 DE 0791751 ES 0791751 FR 0791751 GB 0791751 PT 0791751 AU 93321 ES 107358 FR 222 394 395 PENDING PATENT CATTANI IT M...

Страница 2: ...SE 15 INSTALLATION 16 MOUNTING 17 NOTICES 17 GENERAL WARNINGS AND BIOLOGICAL RISK 17 ROUTINE MAINTENANCE 18 FAULTS REPORT AND POSSIBILE SOLUTIONS 18 EXTRAORDINARY MAINTENANCE AND ROUTINE ANNUAL FIVE Y...

Страница 3: ...12...

Страница 4: ...or their functioning must never be modified Field of use for the German market only According to the annex no 50 of the General Framework of the Administrative Standard regarding the minimum requireme...

Страница 5: ...will find some details listed below SEPARATORE CENTRIFUGO COMPATTO Name SEPARATORE CENTRIFUGO COMPATTO Electric power supply 230 240 V 1 50 Hz 2 8 A 230 400 V 3 50 Hz 2 2 1 3 A Maximum allowed negativ...

Страница 6: ...won t be possible to start the machine again unless the amalgam collection container 5 is replaced Every time an alarm causes the machine to stop a volt free clean contact closes terminals 25 26 draw...

Страница 7: ...since the contained material is toxic and contaminating for epidemic diseases we recommend to handle it with care To receive a list of authorized companies with licence for transport and disposal of t...

Страница 8: ...e installer should train the surgery staff with practical demonstra tions for amalgam container replacement and filter cleaning on new and therefore not contaminated machines Installation connection m...

Страница 9: ...liquid leaks In case of working anomalies or alarms contact the technician Faults report and possibile solutions The amalgam container filling levels are the only alarms directly linked to the Hydro...

Страница 10: ...15 inserting a metallic piece between the two photodiodes The Hydrocyclone control panel or the display will show a visual alarm to indicate that the amalgam level has reached 95 or 100 3 Remove and...

Страница 11: ...tant Remove the liquids and residual debris which are inside the appliance place it in a poly ethylene bag seal and pack it in packaging constructed of 3 layer corrugated board Important notices The m...

Страница 12: ...l ssigkeit Ricircolo Re circulation Umlauf Tubo in ingresso separatore Separator hose inlet Abscheidervorgang Uscita liquido depurato Processed liquid outlet Ablauf der gereinigten Fl ssigkeit Illustr...

Страница 13: ...Fig 2 Draw 2 Abb 2 SEPARATORE D AMALGAMA IDROCICLONE ISO 18 MONTAGGIO AMALGAM SEPARATOR HYDROCYCLONE ISO 18 MOUNTING AMALGAMABSCHEIDER HYDROZYKLON ISO 18 MONTAGE 42...

Страница 14: ...MOUNTING AMALGAMABSCHEIDER HYDROZYKLON ISO 18 MONTAGE all aspiratore to the aspirator zur Saugmaschine D 50 mm ai riuniti to the dental units zu den Behandlungseinheiten D 50 mm Separatore aria acqua...

Страница 15: ...Fig 4 Draw 4 Abb 4 SEPARATORE D AMALGAMA IDROCICLONE ISO 18 MONTAGGIO AMALGAM SEPARATOR HYDROCYCLONE ISO 18 MOUNTING AMALGAMABSCHEIDER HYDROZYKLON ISO 18 MONTAGE 44...

Страница 16: ...ISO 18 COLLEGAMENTO AMALGAM SEPARATOR HYDROCYCLONE ISO 18 CONNECTION AMALGAMABSCHEIDER HYDROZYKLON ISO 18 ANSCHLUSS all aspiratore to the aspirator zur Saugmaschine all aspiratore to the aspirator zu...

Страница 17: ...ndicazione d allarme come visualizzata sul display del circuito AC80 Alarm indication as shown on the display of the circuit AC80 Alarmanzeige wie auf dem Display der Platine AC80 angezeigt Report Cod...

Страница 18: ...SEPARATORE D AMALGAMA IDROCICLONE ISO 18 ISTRUZIONI PER L USO AMALGAM SEPARATOR HYDROCYCLONE ISO 18 DIRECTIONS FOR USE AMALGAMABSCHEIDER HYDROZYKLON ISO 18 GEBRAUCHSANWEISUNG Fig 7 Draw 7 Abb 7 47...

Страница 19: ...ROCICLONE ISO 18 MANUTENZIONE AMALGAM SEPARATOR HYDROCYCLONE ISO 18 MAINTENANCE AMALGAMABSCHEIDER HYDROZYKLON ISO 18 WARTUNG Fig 8 Draw 8 Abb 8 Corpo metallico per test Metallic piece for checking Met...

Страница 20: ...SEPARATORE D AMALGAMA IDROCICLONE ISO 18 MANUTENZIONE AMALGAM SEPARATOR HYDROCYCLONE ISO 18 MAINTENANCE AMALGAMABSCHEIDER HYDROZYKLON ISO 18 WARTUNG Fig 9 Draw 9 Abb 9 49...

Страница 21: ......

Страница 22: ...334 DE 0351372 ES 0351372 FR 0351372 GB 0351372 EP 0791751 US 5 779 443 CH 0791751 DE 0791751 ES 0791751 FR 0791751 GB 0791751 PT 0791751 AU 93321 ES 107358 FR 222 394 395 PENDING PATENT CATTANI IT M...

Отзывы: