12
cod: ed. 09-2021
When Blok-Jet Silent is installed on a lower level
than the surgeries, the suction piping must go
down to the level of the suction unit, with a couple
of metres of piping laid horizontally on the same
level before connecting with the suction unit. This
will remove the potential energy that the liquids
produce when flowing downward.
ASSEMBLY