PW-II
Service Manual
26
7.
Error Messages & Solution
7.1.
Beep: Re-zero
The range of “Re-zero” is within
±
2% of full capacity. However, when the scale is out of this
range it automatically beeps to notify to user that “Re-zero” function does not work in this
range.
7.2.
Err: Initial Zero
You may set 2% or 10% of full capacity as Initial Zero.
When you place an item more than the initial zero setting value (2% or 10% of full capacity)
and turn on the scale it display “Err” message and does not work.
However, you can make the scale work if you make the item’s weight be within the initial
zero setting range (2% or 10% of full capacity) by decreasing the value of weight.
7.3.
Init: Failure of Analog Module
“Init” message shows up when there is some problem on analog module.
Please turn off the scale and turn on. If you still have “Init” message on display please
contact CAS A/S service.
7.4.
UNSTA: A/D Value Unstable
The scale displays “UNSTA” when it is in unstable status or in low-power supply.
If the scale is not properly balanced, please adjust the 4 legs at the bottom of the scale (turn
legs clockwise or counterclockwise) so as to center the bubble of the leveling gauge inside the
indicated circle. And if the scale is in low power supply please replace the battery or plug in
adapter.
However, if you still have
“
UNSTA
”
problem please contact CAS A/S service.
Downloaded from
www.Manualslib.com
manuals search engine
Содержание PW-10H
Страница 1: ...PW II Service Manual 1 PW II SERVICE MANUAL Downloaded from www Manualslib com manuals search engine...
Страница 7: ...PW II Service Manual 7 1 5 Sealing Method Downloaded from www Manualslib com manuals search engine...
Страница 17: ...PW II Service Manual 17 4 Exploded View Downloaded from www Manualslib com manuals search engine...
Страница 19: ...PW II Service Manual 19 5 2 Main PCB Rear Downloaded from www Manualslib com manuals search engine...
Страница 20: ...PW II Service Manual 20 5 3 BATTERY PCB Downloaded from www Manualslib com manuals search engine...
Страница 21: ...PW II Service Manual 21 5 4 KEY PCB Downloaded from www Manualslib com manuals search engine...