PW-II
Service Manual
12
2.1.5.
How to confirm SPAN calibrated A/D value (C-1)
(1)
Pressing and holding Calibration Switch please press [POWER] key.
After “CAL” message blinks three times and the version of scale is displayed, “C-0” message
will be displayed.
(2)
Press [TARE] to change “C-0” to “C–1”.
(3)
Press [ZERO] key, and then it displays SPAN calibrated A/D value, “0”.
(4)
When the weighing value of maximum capacity is not stable you may use
[TARE] key to decrease or [HOLD] key to increase the value.
For example, when you weigh maximum capacity on the platter in case of 1/20,000 it
may display “19,999” or “20,001”. You can adjust up or down to “20,000” using [HOLD] or
[TARE] key.
(5)
Press [ZERO] to go to and display “C – 1” and you may scroll it at this point.
Then, if you press [ZERO] key again it automatically saves “Span Factor” on it.
2.1.6.
Back to the normal weighing mode (C-0)
(1)
Under the Calibration Switch ON press [POWER] key.
9
After “CAL” message blinks three times and the version of scale is
displayed, “C-0” message will be displayed.
(2)
You may exit from Calibration mode by pressing [ZERO] key.
Downloaded from
www.Manualslib.com
manuals search engine
Содержание PW-10H
Страница 1: ...PW II Service Manual 1 PW II SERVICE MANUAL Downloaded from www Manualslib com manuals search engine...
Страница 7: ...PW II Service Manual 7 1 5 Sealing Method Downloaded from www Manualslib com manuals search engine...
Страница 17: ...PW II Service Manual 17 4 Exploded View Downloaded from www Manualslib com manuals search engine...
Страница 19: ...PW II Service Manual 19 5 2 Main PCB Rear Downloaded from www Manualslib com manuals search engine...
Страница 20: ...PW II Service Manual 20 5 3 BATTERY PCB Downloaded from www Manualslib com manuals search engine...
Страница 21: ...PW II Service Manual 21 5 4 KEY PCB Downloaded from www Manualslib com manuals search engine...