PW-II
Service Manual
11
2.1.3.
Gravity Constant Value Setting (C-9)
Current gravitational Acceleration value is set to 9.7994 m/s
2
.
(1) Pressing and holding “Calibration Switch” press [POWER] key.
After “CAL” message blinks three times and shows the version of scale, it displays “C-0”
message.
(2) Press [TARE] to display “C-9”.
(3) Press [ZERO] key, and then “ G-1“ message and “9.7994” will be shown. The first
digit,”9” will blink.
(4) Input a gravitational acceleration value by using [HOLD] and [TARE] key.
(5) Press [ZERO] key, and then “G-2“ message blinks.“9.7994” will be shown. The first
digit,”9” will blink.
(6) Input a gravitational acceleration value by using [HOLD] and [TARE] key.
(7) Press [ZERO] key to save the gravitational acceleration value, and “C9 END” message
will be shown shortly.
2.1.4.
Displaying Real A/D Value (C-5)
(1)
Under Calibration switch on press [POWER] key.
After “CAL” message blinks three times and the version of scale is displayed, “C-0” message
will be displayed.
Press [TARE] to display “C-5”.
Press [ZERO] key, and then the display will show a real A/D value.
Press [ZERO] key to exit from it.
Downloaded from
www.Manualslib.com
manuals search engine
Содержание PW-10H
Страница 1: ...PW II Service Manual 1 PW II SERVICE MANUAL Downloaded from www Manualslib com manuals search engine...
Страница 7: ...PW II Service Manual 7 1 5 Sealing Method Downloaded from www Manualslib com manuals search engine...
Страница 17: ...PW II Service Manual 17 4 Exploded View Downloaded from www Manualslib com manuals search engine...
Страница 19: ...PW II Service Manual 19 5 2 Main PCB Rear Downloaded from www Manualslib com manuals search engine...
Страница 20: ...PW II Service Manual 20 5 3 BATTERY PCB Downloaded from www Manualslib com manuals search engine...
Страница 21: ...PW II Service Manual 21 5 4 KEY PCB Downloaded from www Manualslib com manuals search engine...