22
ERROR MESSAGES & TROUBLE SHOOTING
Err 01
■
Cause
The Indicator can't be initialized due to platform shaking
■
Management
Put the platform on a flat place without shaking and power on
If you can see the same massage continuously, call your CAS dealer please.
Err 02
■
Cause
Initial A/D value of zero is too high or low, when the power is ON
Or in the case of NC-I is lying on the flat
■
Management
Check that the load cell is well connected with main unit
Or calibration is done wrongly
Don't lie the NC-1 on the flat
If you can see the same massage continuously, call your CAS dealer please.
Err 09 (Zero Key)
■
Cause
The current weight exceeds zero range (2 % of Maximum Capacity)
■
Management
Unload the scale
If you can see the same massage continuously, call your CAS dealer please.
Err 13 (Zero Range when the Power is ON)
■
Cause
The initial zero range exceeds above 10% of maximum capacity
■
Management
Unload the scale
If you can see the same massage continuously, call your CAS dealer please.