67
13. OPTIONS
OP-1
Analog Output Interface (Current Output)
■
Set F56 to 1 in set mode.
■
Specification
Output Current
4 – 20mA, 0 – 20mA, 0 – 24mA
Resolution
More than 1/2000
Temperature Coefficient
0.01%
Maximum Load Resistor
500
Ω
MAX.
■
When the display weight is "0", the output current is 0mA or 4mA.
When the display weight is maximum capacity of the indicator, the output current is
20mA or 24mA.
■
To use current to voltage
If you add a 250
Ω
shunt resistor, voltage output will be 1 - 5V or 0 – 5V or 0 – 6V.
(250
Ω
* 4mA
∼
250
Ω
* 20mA)
Shunt Resistor
+ 15V
HI
LO