18
CAL 3
FUNCTION : Setting Weight In Span CALIBRATION
KEY
DISPLAY
DESCRIPTION
L=05000
5000 kg
:
Increase or decrease
of number
:
Shift of cursor position
ENTER :
Save and go to next
menu
L=0500.0
500.0 kg (First decimal point)
Note 1.
The setting weight should be within the 10 % to 100 % of maximum weight
100% of maximum weight is set as a default but you can change it for your needs
If the setting weight is under 10%, error message "Err 22" is displayed
If the setting weight is over 100%, error message "Err 23" is displayed
CAL 4
FUNCTION : Zero Calibration
KEY
DISPLAY
DESCRIPTION
Remove an item from the platter
and press ENTER key.
Zero calibration.
ENTER :
Zero calibration and
next
ZERO :
Only zero calibration
TARE :
Only span calibration
CAL 4
Analog value of
load weight state
- - -
Zero calibration is finished.
You will go to span
calibration(CAL 5)automatically.
Note 1. If Zero calibration is done without any error, You will go to span calibration (CAL 5)
automatically.
Note 2. If the zero value is too low, error message "Err 27" is displayed.
Note 3. If the zero value is too high, error message "Err 26" is displayed.
Note 4. If you want to do only zero calibration, unload the platter and press the "ZERO" key.
In a few moment, you will see "ZEro" and "CAL End" messages are displayed.
Set SW1 DIP1 to OFF and then return to the weighing mode.
Содержание CI-6000A Series
Страница 1: ...2 ...
Страница 8: ...9 4 Measure of Appearance ...
Страница 22: ...23 5 Sealing Method Sealing method of cal switch Sealing method of Load Cell Connector ...
Страница 51: ...52 Normal Batching Diagram Loss in Weight Batching Diagram ...
Страница 52: ...53 USER PROGRAM CONTROLMODE Normal Batching ...
Страница 54: ...55 USER PROGRAM CONTROLMODE Loss in Weight ...
Страница 56: ...57 AUTOMATIC PROGRAM CONTROLMODE Normal Batching ...
Страница 58: ...59 AUTOMATIC PROGRAM CONTROLMODE Loss in Weight ...
Страница 71: ...72 MEMO ...
Страница 72: ...73 MEMO ...
Страница 73: ...74 MEMO ...
Страница 74: ...75 ...