Note: the chamber’s light intensity specification is with a distance of 6 inches (15 cm)
beneath the lamps in the center of the shelf. It is expectant that lower light readings will
be obtained further away from the lights and at the edge of the shelf.
A positive value will move the light intensity ‘up’ and a negative value ‘down’. Press the
‘home’ button and verify the proper light intensity is displayed. (See Temperature
Calibration Example.)
Calibrating Optional Chart Recorders
For calibrating the optional front and side mounted chart recorders, refer to section
(Optional Accessory Operation)
Light sensor calibration
must be
performed with the
lights ‘on’ and in direct lighting. Caron recommends
calibrating the lights
after
the temperature &
humidity (if applicable) are calibrated.
Содержание 7310-22
Страница 1: ......
Страница 115: ...7310 Series Operations Manual REV E 11 19 19 Page 115 of 131 ELECTRICAL SCHEMATICS ...
Страница 116: ...7310 Series Operations Manual REV E 11 19 19 Page 116 of 131 ...
Страница 117: ...7310 Series Operations Manual REV E 11 19 19 Page 117 of 131 ...
Страница 118: ...7310 Series Operations Manual REV E 11 19 19 Page 118 of 131 ...
Страница 119: ...7310 Series Operations Manual REV E 11 19 19 Page 119 of 131 ...
Страница 120: ...7310 Series Operations Manual REV E 11 19 19 Page 120 of 131 ...
Страница 121: ...7310 Series Operations Manual REV E 11 19 19 Page 121 of 131 ...
Страница 122: ...7310 Series Operations Manual REV E 11 19 19 Page 122 of 131 ...
Страница 123: ...7310 Series Operations Manual REV E 11 19 19 Page 123 of 131 ...
Страница 124: ...7310 Series Operations Manual REV E 11 19 19 Page 124 of 131 ...