Do not use water that contains chloramines.
Chloramines can damage internal rubber
gaskets resulting in leaks.
Use only distilled or deionized water with a
resistivity between 50K
Ω
-CM and 1M
Ω
-CM and
a pH of greater than 6.5. Using water outside
this range will void your warranty.
Connecting the Water Supply (HUMD312, HUMD313 & HUMD314)
HUMD312, HUMD313, HUMD314 control humidity from ambient to 85%,and dehumidify
with ambient air. To ensure proper operation, distilled or deionized water is required as a
supply on units that have humidity control. If these water sources are not available
contact Caron customer service.
CO
2
sample
port
CO
2
supply
inlet