38 Chapter 5 Acquiring Panoramic Images
Launching the X-ray
To launch the X-ray, follow these steps:
1.
Ask the patient to remain still, eyes closed, mouth closed, and tongue placed on the palate.
Leave the X-ray room and close the door. You must keep visual contact with the patient during
acquisition.
2.
Launch the X-ray with the remote control to acquire an image
.
The
turns yellow and a warning sound is heard, indicating X-ray emission.
Press and hold the exposure button until the end of the acquisition is indicated by the on-screen
“Release Handswitch” message.
When the acquisition ends, the
Acquisition
interface disappears and the acquired image is
automatically transferred to the
Imaging window
.
3.
Ask the patient to stay in the same position but with mouth opened. Repeat Step 2 to acquire
another image with the mouth opened.
Important: To stop the acquisition, if there is any problem, release
the exposure button of the remote control or press the red
emergency stop button.
Содержание CS 8100 3D
Страница 1: ...User Guide CS 8100 3D Family CS 8100 3D CS 8100 3D Access CS 8100SC 3D CS 8100SC 3D Access...
Страница 6: ...vi...
Страница 8: ...2 Chapter 1 Conventions in This Guide...
Страница 24: ...18 Chapter 3 Imaging Software Overview Note Optional for CS 8100 3D Access and CS 8100SC 3D Access...
Страница 28: ...22 Chapter 3 Imaging Software Overview...
Страница 32: ...26 Chapter 4 Getting Started...
Страница 46: ...40 Chapter 5 Acquiring Panoramic Images...
Страница 72: ...66 Chapter 7 3D Object Acquisition The completely reconstructed object will display in the image browser...
Страница 94: ...88 Chapter 8 Maintenance...
Страница 96: ...90 Chapter 9 Troubleshooting...