Page 50 of 52
RD-OI-48 A/0
and
70% U
T
; 25/30 cycles
Single phase: at 0º
Voltage
interruptions
IEC 61000-4-11
0% U
T;
250/300 cycles
Table 25 Signal input/output parts Port
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility environment
Electrical fast
transients/burst
IEC 61000-4-4
±1 kV
100kHz repetition frequency
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and 80MHz
80%AM at 1kHz
12
Abbreviations
Table 26 Abbreviations
Abbreviation
Description
OEM
Original equipment manufacturer
SIP/SOP
Signal input/output part
FCC
Federal communications committee
EMI
Electromagnetic interference
EMS
Electromagnetic susceptibility
EMC
Electromagnetic compatibility
IEEE
Institute of electrical and electronics engineers
API
Application program interface
DR
Digital radiography
DROC
Digital radiography operator console
F
2
AED
Full-Field automatic exposure detection
TFT/PD
Thin-film transistor/ Photo-diode
DQE
Detective quantum efficiency