Page 47 of 52
RD-OI-48 A/0
Interface cable
7.0 m
Shielded
1 Set
DC Power
Gigabit LAN cable
10 m
Shielded
1 Set
-
Important information regarding Electro Magnetic Compatibility (EMC)
X-ray flat panel detectors needs special precautions regarding EMC and put into service according to
the EMC information provided in the operation manual; X-ray flat panel detectors conform to this IEC
60601-1-2:2014 standard for both immunity and emissions. Nevertheless, special precautions need to
be observed:
X-ray flat panel detectors following essential performance is intended used in professional
healthcare facility environment.
Essential performance: Image acquisition has no negative effects and Image data transmission has
no negative effects
WARNING: Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the
X-ray flat panel detectors, including cables specified by the manufacturer. Otherwise, degradation
of the performance of this equipment could result.”
The use of accessories and cables other than those specified by CareRay, with the exception of
accessories and cables sold by CareRay of X-ray flat panel detectors as replacement parts for
internal components, may result in increased EMISSIONS or decreased IMMUNITY of the X-ray
flat panel detectors.
WARNING: Use of this equipment X-ray flat panel detectors adjacent to or stacked with other
equipment should be avoided because it could result in improper operation.”
When the AC input voltage is interrupted, the X-ray flat panel detectors will power off and if the
power supply restored, it should be recovered by operator manually, this degradation could be
accepted because it will not lead to unacceptable risks and it will not result in the loss of basic
safety or essential performance
EMI Compliance Table
Table 21 Emission
Phenomenon
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1, Class A
Professional healthcare facility environment