Page 19 of 52
RD-OI-48 A/0
Note
: The detector can’t terminate x-ray exposure via communication with the generator.
3.2.1
Software Sync Mode
The detector communicates with an X-ray generator via the DROC workstation via a wired or wireless
network connection. Compared with the external sync mode, the software sync mode will cause a
delayed response.
The sequence chart of software sync mode for a complete cycle of image acquisition is shown in figure
below.
Hand Switch 2
nd
Gear
X-ray ON
pressed
Exposure Time
released
SOFT_SYNC_EXP_REQ
SOFT_SYNC_EXP_EN
Figure 7 Sequence chart
A typical configuration for software sync mode of a DR system integrated with CareView 1500Cwe is
shown in the figure below.
X-ray Tube
X-ray Generator
DROC Workstation
Shielded X-ray Room
RS232 Cable
Gigabit LAN Cable
Wireless AP
CareView 1500Cwe
IEEE802.11n
Control Panel
Figure 8 Software sync mode
3.2.2
Manual Sync Mode
There is no sync exposure signal interaction between the detector and an X-ray generator. Users should
control exposure time manually according to exposure recommendations provided by acquisition