Page 38 of 52
RD-OI-48 A/0
8)
Before discarding the battery, cover the terminals with adhesive tape or other insulators.
Contact with other metal materials may cause fire or explosion.
Calibration
Image quality will degrade due to X-ray exposure and component degradation. Therefore, the detector
must be calibrated for initial use and every following three month interval.
7.3
Cleaning
External cleaning
Use a dry cloth to clean the surface of the product. Don’t use strong detergents or organic solvents to
clean the product. Detergents and organic solvents may damage the product and cause structural
weakening. The product must be turned off during cleaning.
Internal Cleaning
The CareView 1500Cwe has a completely enclosed design which eliminates the need for internal
cleaning.
Note:
Do not use an abrasive brush, scraper, or acid/alkaline cleaner when cleaning the product.
7.4
Emergency
In the event of any of the below situations, immediately turn OFF the power of each device and unplug
the power cable from the AC outlet. In case of an accident, guide the patient far away from the X-ray
exposure area.
Any part of the system emits smoke or an irregular smell or sound.
The X-ray continues to expose or not expose at all during the diagnostic process.
The detector reaches an excessively high temperature.
The detector is dropped and damaged.
7.5
Maintenance Note
Do not open and retrofit the detector as this will invalidate the guarantee of product quality and service.
After-sales service for our product will be provided within warranty period. We will not be responsible
for damage to the product due to incorrect or unsafe handling of the product.
8
Storage and Transportation
8.1
Storage
Store in a dry and well-ventilated area, and away from strong sunlight and corrosive gas. Do not invert
or heap up more than three layers. Do not store more than 100kg of materials on top of the detector.