RD-OI-48 A/0
X-ray exposure value of 130μGy.
X-ray source position
The angle between the X-ray central propagation axis and the normal axis of the front-end
entrance plane taken at the centre point of the matrix should remain within the following range:
X-ray angle: Min = -45 deg
X-ray angle: Typ = 0 deg
X-ray angle: Max = +45 deg
The distance between the X-ray source focal spot and the front-end entrance plane (SID) should
remain within the following range:
SID: Min = 100 cm
SID: Typ = 135 cm
SID: Max = 180 cm
X-ray window duration
While the X-ray window duration can be functionally settled by the system over a wide range,
image quality can be directly impacted by X-ray window duration (due to leakage, noise, or lag).
The full performance of the detector is guaranteed for the following X-ray window range:
X-ray window duration: Min = 10 ms (TBC)
X-ray window duration: Typ = 500 ms (TBC)
X-ray window duration: Max = 2500 ms (TBC)
3.1.3
Workstation (Recommended)
Table 7 Workstation
Item
Description
Operating system
Windows XP/Win7/Win10
Processor
Intel i5 or compatible with AMD processor
RAM
>
6GByte
HDD
>
80GByte
Network
Dual 1000Mbit
3.2
X-ray Sync Mode
Synchronous connection mode is the signal transfer mode between the X-ray generator, which sends
the X-ray, and the detector, which receives the X-ray. The sync mode of CareRay’s static detector
contains external sync, software sync, manual sync and F
2
AED modes.
CareView 1500Cwe-supported typical sync mode contains software sync, manual sync and F
2
AED
modes.