Infusion Sets

• 

Use only sets dedicated for use with the pump. The use of any other set may cause improper 

operation, resulting in an inaccurate fluid delivery or other potential hazard.

• 

It is recommended that infusion sets are changed according to the instructions in the 'Changing 

the Infusion Set' section.  Carefully read the Directions For Use supplied with the infusion set 

prior to use. 

• 

When combining several apparatus and/or instruments with infusion sets and other tubing, for 

example via a 3-way tap or multiple infusion, the performance of the pump may be affected and 

should be monitored closely.

• 

Uncontrolled flow may result if the infusion set is not properly isolated from the patient i.e. 

closing a tap in the set or activating an in-line clamp / roller clamp.

• 

The infusion set may be fitted with an in-line clamp, which can be used to occlude tubing in case 

it is required to stop fluid flow.

• 

The pump is a positive pressure pump, which should use infusion sets fitted with Luer lock 

fittings or equivalent locking connectors.

• 

To infuse from a burette, close the roller clamp above the burette and open the clamp on the 

vent on top of the burette. 

• 

Before operating the pump, verify the infusion set is free from kinks and installed correctly in 

the pump.

• 

Discard infusion set if the packaging is not intact or the protector cap is detached. Ensure sets 

are not kinked as this may occlude the tubing.

Using Collapsible bags, Glass Bottles and Semi Rigid containers

• 

It is recommended that the air vent be opened on the infusion set if using glass bottles or semi-

rigid containers, to reduce the partial vacuum formed as the fluid is infused from the container. 

This action will ensure the pump can maintain volumetric accuracy whilst the container empties. 

The action of opening the air vent for semi-rigid containers should take place after the spiking of 

the container and priming of the drip chamber.

Steps for Glass Bottles and Semi-rigid containers

4.

5.

3.

2.

Spike the 

container

Fill drip 

chamber to 

fill line

Prime the set by 

opening / closing the 

roller clamp

1.

Close the 

roller clamp

Open the air vent 

to allow pressure 

equalisation - ready 

for infusion

Steps for the Collapsible bags

Follow steps 1 to 3 as shown for the 

semi-rigid containers, however 

do not 

open vent as in step 4, but prime the 

set as per step 5. Ensure the bag outlet  

is fully pierced before filling the drip 

chamber.

Operating Environment

• 

When using any infusion pump in conjunction with other pumps or devices requiring vascular 

access, extra care is advised. Adverse delivery of medication or fluids can be caused by the 

substantial variation in pressures created within the fluid channels of such pumps.  Typical 

examples of those pumps are used during dialysis, bypass or cardiac assist applications.

• 

The pump is suitable for use in all establishments, including domestic establishments and those 

directly connected to the public low voltage power supply network that supplies buildings used 

for domestic purposes.

• 

This pump is not intended to be used in the presence of a flammable anaesthetic mixture with 

air or oxygen or nitrous oxide.

• 

Do not use the pump in a hyperbaric chamber.

• 

Do not use the pump near Magnetic Resonance Imaging (MRI), 

including Stereotaxis technology

.

Operating Precautions

Содержание Alaris SE 7130

Страница 1: ...s Alaris SE Pump Models 7130 7131 and 7230 7231 with Guardrails Safety Software Directions For Use en...

Страница 2: ...Multi Dose 18 Multi Step 19 Monitoring Options 20 SmartSite Needle Free System Instructions 24 Alarms Alerts and Prompts 25 Alarms 25 Alerts 26 Prompts 27 Guardrails Event Counter 29 Flow Sensor Oper...

Страница 3: ...future updates and reference Information about a Guardrails alert that occurs during use is stored within the Pump and can be accessed using the CQI Reporter The Pump may be operated with or without t...

Страница 4: ...SEC KVO OPT 2 3 4 5 6 7 8 9 0 1 POWER RUN HOLD A B A B ml hr ml hr A B A B Latch Pumping Mechanism Pressure Transducer Air in line Detector Air in line Arm Panel Lock Keypad AC power indicator Rate Di...

Страница 5: ...key Silences audible alarm or alert for 2 minutes message remains on screen New alarm or alert reinstates the audible tone PRI Primary key Selects Primary mode Channel must be selected if applicable S...

Страница 6: ...ts of Council Directive 93 42 EEC as amended by 2007 47 EC T Date of Manufacture t Manufacturer B Not for Municipal Waste e Functional Earth RS 232 RS232 Nurse Call Connector Canadian and U S Certific...

Страница 7: ...ing when the pump is turned off Panel lock icon Indicates that panel lock is on Audio Volume Indicator Indicates audio volume for alarms and alerts Low Medium High MNTR Computer Mode Indicator Display...

Страница 8: ...fusion set if using glass bottles or semi rigid containers to reduce the partial vacuum formed as the fluid is infused from the container This action will ensure the pump can maintain volumetric accur...

Страница 9: ...adio interference Reorienting relocating or shielding the system or filtering the connection to the public mains network are examples of steps that can be taken to reduce or eliminate interference How...

Страница 10: ...ibility and pump performance as part of the overall infusion Potential hazards include drug interactions inaccurate delivery rates and pressure alarms and nuisance alarms When delivering an infusion w...

Страница 11: ...use The pump will automatically operate from its internal battery if the pump is switched on without being connected to the power supply Should the pump fail to perform correctly replace in its origi...

Страница 12: ...the manufacturer s directions for use Use only sets dedicated for use with the Alaris SE Pump 1 Slide Flow Regulator thumb clamp down until an audible click verifies it is in fully closed position 2 S...

Страница 13: ...ow Regulator A After set installation verify no fluid is flowing through the infusion set s drip chamber to avoid free flow Before operating pump verify that infusion set is free from kinks and instal...

Страница 14: ...ress RUN HOLD key to start infusion 5 To briefly view current profile from RUN HOLD page or during a running infusion press 4 soft key New Guardrails Drug 1 Select a drug name by pressing the adjacent...

Страница 15: ...oft key 3 Press soft key next to parameter to be edited Current value is highlighted 4 To enter a new value use numeric keypad then press ENTER key Getting Started continued A Rate LED flashes while i...

Страница 16: ...guration settings for the selected profile and the dosing limits for the selected drug to be viewed 1 Press OPTIONS key Options page appears 2 To view additional settings press page soft key Second op...

Страница 17: ...nfusion set 6 Press the SEC key Secondary Setup page displays 7 To enter desired Rate use numeric keypad then press ENTER key 8 To enter desired VTBI use numeric keypad then press ENTER key 9 Verify t...

Страница 18: ...y or to enter a new vtbi use numeric keypad and press ENTER key 6 To start loading dose infusion press RUN HOLD key 7 To briefly view current profile press 4 soft key A This mode is useful for deliver...

Страница 19: ...Alert Option is enabled DOSE COMPLETE ALERT OPTION page appears To select On or Off use soft keys 11 To continue programming press ok soft key 12 To start first dose immediately press ok soft key The...

Страница 20: ...e channel on hold press RUN HOLD key 2 To return to review page s press setup soft key A tick mark y next to a step on review page s indicates it has not started Only steps having a y can be edited A...

Страница 21: ...stable Pressure designed to monitor IV infusion set site pressure and provide user adjustable pressure alarm limits Used for Precision Flow mode or for high resistance systems such as infusion through...

Страница 22: ...pressure limit is exceeded This limit may be set by qualified service personnel from 1 mmHg to 600 mmHg Pressure Limit Maximum Resistance Alert The Resistance Alert provides an early warning of increa...

Страница 23: ...timal baseline is maintained upon subsequent activations of RUN HOLD as follows If current system pressure is same or higher than original baseline pressure baseline does not change If current system...

Страница 24: ...alibrate based on current system pressure 1 To place channel on hold press channel s RUN HOLD key All infusions connected to channel being baselined must be on hold 2 Press options key 3 Press Set Pre...

Страница 25: ...S Use Aseptic Technique 1 Prior to every access swab top of Needle Free Valve port with 70 Isopropyl alcohol 1 2 seconds and allow to dry approximately 30 seconds NOTE Dry time is dependent on tempera...

Страница 26: ...ual channel pump only Channel malfunction Turn channel off and then on If problem persists do not use channel Contact qualified service personnel computer link failure RS 232 connection to computer wa...

Страница 27: ...nked tubing closed clamp Verify that blue thumb clamp on Flow Regulator has moved to open up position If not reload set Press run soft key to restart infusion primary flow detected during secondary Pu...

Страница 28: ...ode 600 ml h in secondary mode Verify and re enter settings New Baseline Set A new Manual Pressure Baseline was successfully set Manual Pressure Baseline feature is on Baseline remains set until a new...

Страница 29: ...TBI or lower Loading Dose VTBI as appropriate Stop Timer to Change An invalid key was pressed while timer was running in Multi Dose program Wait several seconds for popup to finish Press stop timer so...

Страница 30: ...Events maximum rate maximum weight soft dose limit high or low hard dose limit high or low drug selection cancelled Overrides Total number of Guardrails Soft Limits that are overridden Counter increme...

Страница 31: ...in a No Upstream Flow Detected alarm This is because the small rapidly falling drops form a continuous stream which does not trigger the flow sensor In this event unplug the flow sensor from the pump...

Страница 32: ...arts If allowable number of restarts is exceeded or when resistance or pressure conditions indicate an occlusion an occlusion downstream alarm occurs Monitoring Options Dynamic monitoring provides cli...

Страница 33: ...Library Bolus Dose feature may be used Multi Dose Allows 1 24 doses to be programmed at equally spaced intervals on same pump over a 24 hour period This mode allows delivery of multiple equal doses fr...

Страница 34: ...lus Off On Off Dose Rate Calculator Off On Off Loading Dose On On Off Multi Dose Off On Off Multi Dose Alert Off On Off Multi Step Off On Off Panel Lock On On Off Rate Maximum 999 9 ml h 0 1 999 9 ml...

Страница 35: ...oint failure does not exceed 1 0 ml at 999 9 ml h Dimensions Nominal 7131 7231 Depth 127mm 127mm Height 218mm 218mm Width 193mm 267mm Weight 3 0 kg 3 8 kg Without pole clamp Without power cord Downstr...

Страница 36: ...listed below Rate Accuracy Test Conditions Infusion rate range 0 1 to 999 9 ml h Head height 24 1 in 61 2 5 cm Test solution distilled water Environment temperature 68 8 F 20 4 C Back pressure 0 psi N...

Страница 37: ...accuracy is reached after several complete charge discharge recharge cycles in the refresh process It is recommended that the battery be fully charged discharged recharged using the refresh cycle bef...

Страница 38: ...y running warm water over it while actuating slider and then thoroughly dried After cleaning the sensor should be allowed to dry fully prior to use Disposal Information on Disposal for Users of Waste...

Страница 39: ...ith the cable jacket Connecting to a Computer 1 Press options key 2 To advance to next page press page soft key 3 Press Computer Link soft key 4 Press Monitor soft key 5 Press ok soft key 6 Connect an...

Страница 40: ...also be considered The start up curves represent continuous flow rate versus operating time for two hours from the start of the infusion They exhibit the delay in onset of delivery due to mechanical c...

Страница 41: ...at 1 mL h initial 0 8 1 1 4 9 4 9 6 2 2 0 12 1 3 9 21 2 12 2 3 2 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Pressure Mode Trumpet Curve at 0 1 mL h 48 hr 21 8 11 9...

Страница 42: ...Pressure Mode Trumpet Curve at 999 9 mL h initial 1 4 1 3 0 9 1 0 1 1 1 1 1 2 1 8 1 3 1 5 1 2 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Pressure Mode Trumpet Curve...

Страница 43: ...4 8 3 7 11 5 0 6 23 4 2 6 32 6 13 3 1 2 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Resistance Mode Trumpet Curve at 0 1 mL h 48 hr 22 9 22 0 46 4 17 2 48 9 6 7 109...

Страница 44: ...erval min Flow Rate Error Resistance Mode Trumpet Curve at 999 9 mL h 24 hr 3 5 3 4 2 7 2 9 3 1 3 1 3 9 3 3 3 7 3 2 3 3 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Hi...

Страница 45: ...mark CareFusion Via Ticino 4 50019 Sesto Fiorentino Firenze Italia CareFusion Hammarbacken 4B 191 46 Sollentuna Sverige Tel 32 2 267 38 99 Tlf 45 70 20 30 74 T l 39 055 30 33 93 00 Tel 46 8 544 43 200...

Страница 46: ...on This Class A b digital apparatus meets all requirements of the Canadian Interference Causing Equipment Regulation Tables The Pump is intended for use in the electromagnetic environments specified i...

Страница 47: ...y should be that of a typical commercial or hospital environment IEC EN 61000 4 8 Power Frequency Magnetic Field 50 60 Hz 3 A m 400 A m 50 Hz Note 2 Power frequency magnetic fields should be at levels...

Страница 48: ...arked with the following symbol Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absor...

Страница 49: ...2 00 23 00 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applicable to the frequency of the t...

Страница 50: ...ument contains proprietary information of CareFusion Corporation or one of its subsidiaries and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell...

Отзывы: