1000DF00480 Issue 2 29/48
Alarms, Alerts and Prompts (continued)
Guardrails® Event Counter
The pump keeps a running count of Guardrails® Events that occur that can be viewed with the Guardrails® Events counter
feature.
1. Press
OPTIONS
key.
2. Press
Event Counters
soft key.
3. To return to main display, press
return
soft key. To clear counter (reset all values to zero), press
clear
soft key. To view
second event counter page, press
page
soft key.
Definitions:
Events
Total number of Guardrails® Events: maximum rate, maximum weight, soft dose limit (high or low), hard dose limit (high
or low), drug selection cancelled.
Overrides
Total number of Guardrails® Soft Limits that are overridden. Counter increments when
yes
is selected in response to soft
override alert prompt.
Soft Alerts
Total number of Guardrails® Soft Limits not overridden. Counter increments when
no
is selected in response to soft
override alert prompt.
Starts
Total number of infusion starts. A start occurs when
run
key is selected from an inactive or paused condition.
A
•
For dual channel pumps, the counters represent the total counts for both channels.
•
These are the counter labels as they appear on the Main Display.