CareFusion Alaris SE 7130 Скачать руководство пользователя страница 24

1000DF00480 Issue 2                 23/48               

Monitoring Options (continued)

Options (continued)

Pressure Baseline

The Pressure Baseline feature, when enabled through the hospital data set profile configuration settings, provides a real-time bar 

graph and numeric display of line pressure.
The pressure limit may be reduced if the pressure in the set is high or changing.  This results in the pressure limit being lowered 

from the selected setting.  If this occurs, first try to remove or reduce the downstream pressure.  Following that, try to reload the 

set, wait 15 to 30 seconds and then perform a Set Pressure Baseline operation.  The pressure baseline may need to be set a second 

time, after the pressure readings have stabilized.
For dual channel pumps, select the desired channel as necessary.  The pressure bar graph is not shown when the split screen 

display is active.
For optimal results, set the baseline 15 minutes after starting an infusion.  The pressure baseline can be optimised, particularly 

at low flow rates (less than 3 ml/h), by resetting the pressure baseline when the readings are negative.  Check periodically for 

negative readings; for example, when programming VTBI.  This allows the pressure baseline to calibrate based on current system 

pressure.
1.  To place channel on hold, press channel’s 

RUN/HOLD

 key.  (All infusions connected to channel being baselined must be 

on hold.)

2.  Press 

OPTIONS

 key.

3.  Press 

Set Pressure Baseline

 soft key.

4.  Verify no pressure, due to occlusion or other infusions through a common set, is present in infusion set at this time.

5.  For best results, verify set outlet (for example, stopcock) is located at patient’s heart level before continuing with next step.

6.  Press 

ok

 soft key.

7.  Verify pressure readout is 0 (zero) mmHg.

8.  To start infusion, press 

RUN/HOLD

 key. 

A

• 

To return to the normal screen without setting the baseline, press return soft key.

• 

True baseline pressure will be zero or within a few mmHg of zero.  If not, and the pressure is 

unstable, allow the pressure to drop to the lowest level and then repeat the Set Pressure Baseline 

process.

• 

The pressure baseline calibration remains in effect until the pump is turned off, the latch is opened, 

the set is reloaded, or the set Pressure Baseline function is performed again.

• 

Setting the manual baseline overrides the auto baseline until the pump is turned off, the latch is 

opened, set is loaded, or another manual baseline is set.

• 

Setting a manual Pressure Baseline displays a horizontal real-time bar graph and numeric pressure 

readings.  The vertical line on the pressure bar graph visually indicates the pressure alarm limit.

Содержание Alaris SE 7130

Страница 1: ...s Alaris SE Pump Models 7130 7131 and 7230 7231 with Guardrails Safety Software Directions For Use en...

Страница 2: ...Multi Dose 18 Multi Step 19 Monitoring Options 20 SmartSite Needle Free System Instructions 24 Alarms Alerts and Prompts 25 Alarms 25 Alerts 26 Prompts 27 Guardrails Event Counter 29 Flow Sensor Oper...

Страница 3: ...future updates and reference Information about a Guardrails alert that occurs during use is stored within the Pump and can be accessed using the CQI Reporter The Pump may be operated with or without t...

Страница 4: ...SEC KVO OPT 2 3 4 5 6 7 8 9 0 1 POWER RUN HOLD A B A B ml hr ml hr A B A B Latch Pumping Mechanism Pressure Transducer Air in line Detector Air in line Arm Panel Lock Keypad AC power indicator Rate Di...

Страница 5: ...key Silences audible alarm or alert for 2 minutes message remains on screen New alarm or alert reinstates the audible tone PRI Primary key Selects Primary mode Channel must be selected if applicable S...

Страница 6: ...ts of Council Directive 93 42 EEC as amended by 2007 47 EC T Date of Manufacture t Manufacturer B Not for Municipal Waste e Functional Earth RS 232 RS232 Nurse Call Connector Canadian and U S Certific...

Страница 7: ...ing when the pump is turned off Panel lock icon Indicates that panel lock is on Audio Volume Indicator Indicates audio volume for alarms and alerts Low Medium High MNTR Computer Mode Indicator Display...

Страница 8: ...fusion set if using glass bottles or semi rigid containers to reduce the partial vacuum formed as the fluid is infused from the container This action will ensure the pump can maintain volumetric accur...

Страница 9: ...adio interference Reorienting relocating or shielding the system or filtering the connection to the public mains network are examples of steps that can be taken to reduce or eliminate interference How...

Страница 10: ...ibility and pump performance as part of the overall infusion Potential hazards include drug interactions inaccurate delivery rates and pressure alarms and nuisance alarms When delivering an infusion w...

Страница 11: ...use The pump will automatically operate from its internal battery if the pump is switched on without being connected to the power supply Should the pump fail to perform correctly replace in its origi...

Страница 12: ...the manufacturer s directions for use Use only sets dedicated for use with the Alaris SE Pump 1 Slide Flow Regulator thumb clamp down until an audible click verifies it is in fully closed position 2 S...

Страница 13: ...ow Regulator A After set installation verify no fluid is flowing through the infusion set s drip chamber to avoid free flow Before operating pump verify that infusion set is free from kinks and instal...

Страница 14: ...ress RUN HOLD key to start infusion 5 To briefly view current profile from RUN HOLD page or during a running infusion press 4 soft key New Guardrails Drug 1 Select a drug name by pressing the adjacent...

Страница 15: ...oft key 3 Press soft key next to parameter to be edited Current value is highlighted 4 To enter a new value use numeric keypad then press ENTER key Getting Started continued A Rate LED flashes while i...

Страница 16: ...guration settings for the selected profile and the dosing limits for the selected drug to be viewed 1 Press OPTIONS key Options page appears 2 To view additional settings press page soft key Second op...

Страница 17: ...nfusion set 6 Press the SEC key Secondary Setup page displays 7 To enter desired Rate use numeric keypad then press ENTER key 8 To enter desired VTBI use numeric keypad then press ENTER key 9 Verify t...

Страница 18: ...y or to enter a new vtbi use numeric keypad and press ENTER key 6 To start loading dose infusion press RUN HOLD key 7 To briefly view current profile press 4 soft key A This mode is useful for deliver...

Страница 19: ...Alert Option is enabled DOSE COMPLETE ALERT OPTION page appears To select On or Off use soft keys 11 To continue programming press ok soft key 12 To start first dose immediately press ok soft key The...

Страница 20: ...e channel on hold press RUN HOLD key 2 To return to review page s press setup soft key A tick mark y next to a step on review page s indicates it has not started Only steps having a y can be edited A...

Страница 21: ...stable Pressure designed to monitor IV infusion set site pressure and provide user adjustable pressure alarm limits Used for Precision Flow mode or for high resistance systems such as infusion through...

Страница 22: ...pressure limit is exceeded This limit may be set by qualified service personnel from 1 mmHg to 600 mmHg Pressure Limit Maximum Resistance Alert The Resistance Alert provides an early warning of increa...

Страница 23: ...timal baseline is maintained upon subsequent activations of RUN HOLD as follows If current system pressure is same or higher than original baseline pressure baseline does not change If current system...

Страница 24: ...alibrate based on current system pressure 1 To place channel on hold press channel s RUN HOLD key All infusions connected to channel being baselined must be on hold 2 Press options key 3 Press Set Pre...

Страница 25: ...S Use Aseptic Technique 1 Prior to every access swab top of Needle Free Valve port with 70 Isopropyl alcohol 1 2 seconds and allow to dry approximately 30 seconds NOTE Dry time is dependent on tempera...

Страница 26: ...ual channel pump only Channel malfunction Turn channel off and then on If problem persists do not use channel Contact qualified service personnel computer link failure RS 232 connection to computer wa...

Страница 27: ...nked tubing closed clamp Verify that blue thumb clamp on Flow Regulator has moved to open up position If not reload set Press run soft key to restart infusion primary flow detected during secondary Pu...

Страница 28: ...ode 600 ml h in secondary mode Verify and re enter settings New Baseline Set A new Manual Pressure Baseline was successfully set Manual Pressure Baseline feature is on Baseline remains set until a new...

Страница 29: ...TBI or lower Loading Dose VTBI as appropriate Stop Timer to Change An invalid key was pressed while timer was running in Multi Dose program Wait several seconds for popup to finish Press stop timer so...

Страница 30: ...Events maximum rate maximum weight soft dose limit high or low hard dose limit high or low drug selection cancelled Overrides Total number of Guardrails Soft Limits that are overridden Counter increme...

Страница 31: ...in a No Upstream Flow Detected alarm This is because the small rapidly falling drops form a continuous stream which does not trigger the flow sensor In this event unplug the flow sensor from the pump...

Страница 32: ...arts If allowable number of restarts is exceeded or when resistance or pressure conditions indicate an occlusion an occlusion downstream alarm occurs Monitoring Options Dynamic monitoring provides cli...

Страница 33: ...Library Bolus Dose feature may be used Multi Dose Allows 1 24 doses to be programmed at equally spaced intervals on same pump over a 24 hour period This mode allows delivery of multiple equal doses fr...

Страница 34: ...lus Off On Off Dose Rate Calculator Off On Off Loading Dose On On Off Multi Dose Off On Off Multi Dose Alert Off On Off Multi Step Off On Off Panel Lock On On Off Rate Maximum 999 9 ml h 0 1 999 9 ml...

Страница 35: ...oint failure does not exceed 1 0 ml at 999 9 ml h Dimensions Nominal 7131 7231 Depth 127mm 127mm Height 218mm 218mm Width 193mm 267mm Weight 3 0 kg 3 8 kg Without pole clamp Without power cord Downstr...

Страница 36: ...listed below Rate Accuracy Test Conditions Infusion rate range 0 1 to 999 9 ml h Head height 24 1 in 61 2 5 cm Test solution distilled water Environment temperature 68 8 F 20 4 C Back pressure 0 psi N...

Страница 37: ...accuracy is reached after several complete charge discharge recharge cycles in the refresh process It is recommended that the battery be fully charged discharged recharged using the refresh cycle bef...

Страница 38: ...y running warm water over it while actuating slider and then thoroughly dried After cleaning the sensor should be allowed to dry fully prior to use Disposal Information on Disposal for Users of Waste...

Страница 39: ...ith the cable jacket Connecting to a Computer 1 Press options key 2 To advance to next page press page soft key 3 Press Computer Link soft key 4 Press Monitor soft key 5 Press ok soft key 6 Connect an...

Страница 40: ...also be considered The start up curves represent continuous flow rate versus operating time for two hours from the start of the infusion They exhibit the delay in onset of delivery due to mechanical c...

Страница 41: ...at 1 mL h initial 0 8 1 1 4 9 4 9 6 2 2 0 12 1 3 9 21 2 12 2 3 2 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Pressure Mode Trumpet Curve at 0 1 mL h 48 hr 21 8 11 9...

Страница 42: ...Pressure Mode Trumpet Curve at 999 9 mL h initial 1 4 1 3 0 9 1 0 1 1 1 1 1 2 1 8 1 3 1 5 1 2 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Pressure Mode Trumpet Curve...

Страница 43: ...4 8 3 7 11 5 0 6 23 4 2 6 32 6 13 3 1 2 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Resistance Mode Trumpet Curve at 0 1 mL h 48 hr 22 9 22 0 46 4 17 2 48 9 6 7 109...

Страница 44: ...erval min Flow Rate Error Resistance Mode Trumpet Curve at 999 9 mL h 24 hr 3 5 3 4 2 7 2 9 3 1 3 1 3 9 3 3 3 7 3 2 3 3 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Hi...

Страница 45: ...mark CareFusion Via Ticino 4 50019 Sesto Fiorentino Firenze Italia CareFusion Hammarbacken 4B 191 46 Sollentuna Sverige Tel 32 2 267 38 99 Tlf 45 70 20 30 74 T l 39 055 30 33 93 00 Tel 46 8 544 43 200...

Страница 46: ...on This Class A b digital apparatus meets all requirements of the Canadian Interference Causing Equipment Regulation Tables The Pump is intended for use in the electromagnetic environments specified i...

Страница 47: ...y should be that of a typical commercial or hospital environment IEC EN 61000 4 8 Power Frequency Magnetic Field 50 60 Hz 3 A m 400 A m 50 Hz Note 2 Power frequency magnetic fields should be at levels...

Страница 48: ...arked with the following symbol Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absor...

Страница 49: ...2 00 23 00 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applicable to the frequency of the t...

Страница 50: ...ument contains proprietary information of CareFusion Corporation or one of its subsidiaries and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell...

Отзывы: